Farman T T, Farman A G
University of Louisville School of Dentistry, Louisville, Kentucky 40292, USA.
Dentomaxillofac Radiol. 2000 Jan;29(1):41-5. doi: 10.1038/sj/dmfr/4600499.
To compare the properties of a new speed group F intra-oral X-ray film with those of three well established types both from groups D and E, and evaluate the impact of six commercially available processing solutions.
Four types of dental X-ray film, Flow (Flow X-ray, West Hempstead, NY, USA), speed group F, Ektaspeed Plus (Eastman Kodak, Rochester NY, USA) and M2 Comfort (AgfaGevaert, Morstel, Belgium), both speed group E, and Ultra-speed (Eastman Kodak), speed group D, were exposed under standardised conditions and processed in six different processing solutions. Base plus fog density, characteristic curves, film density, speed, average gradient, contrast and latitude and spatial resolution were calculated.
The choice of processing chemistry affected radiographic characteristics including speed grouping. The new F film was consistently the fastest. M2 Comfort could achieve F speed and Ultra-speed achieved E speed using Automat XR chemicals. The speed of Ektaspeed Plus was independent of the automatic processing solution used. Ultra-Speed film had the lowest base plus fog density and the widest latitude. Film contrast was similar irrespective of the film and solution combination.
The choice of processing chemistry can affect radiographic characteristics. The new F film reduces patient exposure by one-half compared with E speed film with no detriment to image quality.
比较新型F速口内X线胶片与D组和E组中三种成熟类型胶片的性能,并评估六种市售冲洗液的影响。
四种牙科X线胶片,即Flow(美国纽约州韦斯特亨普斯特德的Flow X射线公司)、F速组胶片、Ektaspeed Plus(美国纽约州罗切斯特的伊士曼柯达公司)和M2 Comfort(比利时莫斯特尔的爱克发吉华集团),均为E速组胶片,以及Ultra-speed(伊士曼柯达公司)、D速组胶片,在标准化条件下进行曝光,并在六种不同的冲洗液中进行冲洗。计算了片基加灰雾密度、特性曲线、胶片密度、感光度、平均梯度、对比度、宽容度和空间分辨率。
冲洗化学药剂的选择会影响包括感光度分组在内的射线照相特性。新型F胶片始终是最快的。使用Automat XR化学药剂时,M2 Comfort胶片可达到F速,Ultra-speed胶片可达到E速。Ektaspeed Plus胶片的感光度与所使用的自动冲洗液无关。Ultra-Speed胶片的片基加灰雾密度最低,宽容度最宽。无论胶片和冲洗液如何组合,胶片对比度都相似。
冲洗化学药剂的选择会影响射线照相特性。与E速胶片相比,新型F胶片可使患者的辐射剂量减少一半,且不影响图像质量。