Sjöberg F, Nilsson G
Department of Anesthesiology and Intensive Care, Linköping University Hospital, Sweden.
Acta Anaesthesiol Scand. 2000 Jan;44(1):32-6. doi: 10.1034/j.1399-6576.2000.440106.x.
In biomedical research and clinical medicine there is a demand for potent sensors to measure the components that make up blood gas analyses. Today, as when the electrochemical PO2, PCO2 and pH electrodes were first introduced, these measurements are usually made with the same type of sensor technology. The aims of the present study were, firstly, to find out whether the platinum cathode in the Clark electrode can be replaced by antimony for oxygen measurements (amperometry (A)); secondly, whether, during oxygen measurements, the inherent corrosion potential of the antimony metal can be used for measurement of pH in the same measurement area (potentiometry (P)).
An electrode of purified, crystallographically orientated monocrystalline antimony (COMA) connected to a reference electrode (silver-silver chloride) was used for the P measurements. Measurements of A (at -900 mV) and P were made in an aqueous environment regulated for oxygen, pH, and temperature.
Reproducible oxygen sensitivities of 0.925 nA/% oxygen (2% CV (coefficient of variation)) (A), 10.7 mV/% (P), and 0.7 mV/% (P) were found in the oxygen range: 0-21%, <5%, and above 5%, respectively. The pH sensitivity was 57 mV/pH unit (P). Oxygen and pH measurements were less accurate at oxygen concentrations close to 0%.
Both the oxygen and pH part of the composite electrode signal can be identified by this dual mode technique (A and P). The sensor seems to be promising as it provides measurements of two separate variables (oxygen and pH) and also has the desirable characteristics of a solid state sensor.
在生物医学研究和临床医学中,需要强大的传感器来测量构成血气分析的成分。如今,就像当初首次引入电化学PO₂、PCO₂和pH电极时一样,这些测量通常采用相同类型的传感器技术。本研究的目的,首先是要弄清楚克拉克电极中的铂阴极能否被锑取代用于氧气测量(电流分析法(A));其次,在氧气测量过程中,锑金属的固有腐蚀电位能否用于在同一测量区域测量pH(电位分析法(P))。
使用连接到参比电极(银 - 氯化银)的纯化的、晶体取向的单晶锑(COMA)电极进行P测量。在调节了氧气、pH和温度的水性环境中进行A(在 -900 mV)和P测量。
在氧气范围为0 - 21%、<5%和>5%时,分别发现可重现的氧气灵敏度为0.925 nA/%氧气(2%变异系数(CV))(A)、10.7 mV/%(P)和0.7 mV/%(P)。pH灵敏度为57 mV/pH单位(P)。在氧气浓度接近0%时,氧气和pH测量的准确性较低。
通过这种双模式技术(A和P)可以识别复合电极信号中的氧气和pH部分。该传感器似乎很有前景,因为它能提供两个独立变量(氧气和pH)的测量,并且还具有固态传感器的理想特性。