Binda R E, Cook D R, Fischer C G
Anesth Analg. 1976 Nov-Dec;55(6):769-72. doi: 10.1213/00000539-197611000-00004.
A lung simulator with variable compliance and resistance components was used to evaluate the dynamic compliance of the Bournes, Babybird, and Pediatric Emerson postoperative ventilators. With increase in airway pressure from combined changes in compliance and resistance, the internal compliance of the Bournes was lowest and the internal compliance of the Emerson was highest. With low constant airway resistance (50 cm/L/sec), the Babybird exhibited tidal volume losses similar to those of the Bournes in the face of decreased lung compliance. With constant lung compliance (10 ml/cm H2O) and increasing airway resistance, the Babybird had marked volume losses at higher volumes. Under all simulated conditions, internal compliance of the Emerson, although large, was relatively constant and the Bournes had the smallest internal compliance.
使用具有可变顺应性和阻力组件的肺模拟器来评估伯恩斯、小鸟和儿科艾默生术后呼吸机的动态顺应性。随着气道压力因顺应性和阻力的综合变化而增加,伯恩斯的内部顺应性最低,艾默生的内部顺应性最高。在低恒定气道阻力(50 cm/L/秒)的情况下,面对肺顺应性降低,小鸟表现出与伯恩斯相似的潮气量损失。在肺顺应性恒定(10 ml/cm H2O)且气道阻力增加的情况下,小鸟在较高潮气量时出现明显的容量损失。在所有模拟条件下,艾默生的内部顺应性虽然较大,但相对恒定,而伯恩斯的内部顺应性最小。