LCModel在短回波时间STEAM序列的体内1H磁共振波谱质量控制和定量分析中的应用。

Application of LCModel for quality control and quantitative in vivo 1H MR spectroscopy by short echo time STEAM sequence.

作者信息

Hájek M, Burian M, Dezortová M

机构信息

Department of Diagnostic and Interventional Radiology, Institute for Clinical and Experimental Medicine, Prague, Czech Republic.

出版信息

MAGMA. 2000 Feb;10(1):6-17. doi: 10.1007/BF02613107.

Abstract

The linear combination of model spectra (LCModel) calculation of a parameter for long-term quality control, kT, was introduced, representing the ratio of the temporal and nominal intensities of CH3 groups of lactate and acetate in a quality control phantom. This procedure is a part of the quality assurance of the scanner using fully automatic measurement and calculation of kT parameters, and utilizing Shewhart regulation control charts for continuous evaluation of the magnetic resonance (MR) scanner setting. The application of the kT parameter for the correction of in vivo data increases the precision of molar concentration determination by about 4%. This was tested by the quantitative in vivo MR determination of the molar concentrations of 13 prominent metabolites (N-acetylaspartate (NAA), N-acetylaspartylglutamate, creatine and phosphocreatine (Cr), choline-containing compounds (Cho), myo-inositol, scyllo-inositol, gamma-aminobutyric acid, glutamine, glutamate, glucose, lactate, alanine, taurine) in the white matter and hippocampus of the brain in groups of volunteers, using a short echo time stimulated echo acquisition mode sequence (echo time = 10 ms) and the LCModel technique. The repeatability of the measurement of prominent metabolites such as NAA, Cr and Cho was found to be around 10% (relative standard deviation, n = 6); precision in a group of volunteers (n = 20 and 28, respectively) was in the range of approximately 13-20%. For other metabolites, which are measured with a lower signal-to-noise ratio, the precision can be much lower.

摘要

引入了用于长期质量控制的参数kT的模型光谱线性组合(LCModel)计算,它代表质量控制模型中乳酸盐和醋酸盐CH3基团的时间强度与标称强度之比。此程序是扫描仪质量保证的一部分,采用kT参数的全自动测量和计算,并利用休哈特控制图对磁共振(MR)扫描仪设置进行持续评估。kT参数在体内数据校正中的应用使摩尔浓度测定的精度提高了约4%。通过使用短回波时间刺激回波采集模式序列(回波时间 = 10 ms)和LCModel技术,对志愿者组大脑白质和海马体中13种主要代谢物(N - 乙酰天门冬氨酸(NAA)、N - 乙酰天门冬氨酰谷氨酸、肌酸和磷酸肌酸(Cr)、含胆碱化合物(Cho)、肌醇、 scyllo - 肌醇、γ - 氨基丁酸、谷氨酰胺、谷氨酸、葡萄糖、乳酸盐、丙氨酸、牛磺酸)的摩尔浓度进行定量体内MR测定对此进行了测试。发现NAA、Cr和Cho等主要代谢物测量的重复性约为10%(相对标准偏差,n = 6);一组志愿者(分别为n = 20和28)的精度在约13 - 20%范围内。对于其他以较低信噪比测量的代谢物,精度可能会低得多。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验