Wang Y, Fice D S, Yeung P K
Pharmacokinetics and Metabolism Laboratory, College of Pharmacy, Burbridge Building, Dalhousie University, Halifax, Canada.
J Pharm Biomed Anal. 1999 Nov;21(3):519-25. doi: 10.1016/s0731-7085(99)00117-x.
A reversed-phase HPLC assay coupled with electrochemical detection for simultaneously measuring plasma levels of norepinephrine, epinephrine, dopamine, and 3,4-dihydroxyphenylacetic acid (DOPAC) was developed. Separation of the catecholamines and the internal standard isoproterenol was obtained by a mobile phase consisting of 7% methanol in 0.1 M citrate buffer containing 0.3 mM sodium ethylenediaminetetraacetic acid (EDTA), and 0.5 mM 1-octanesulfonic acid, operated under isocratic condition at a flow rate of 1.2 ml/min. The potential of the guard cell was set at +650 mV, the first electrode of the analytical cell at +100 mV and the second at + 350 mV. Using a signal-to-noise ratio of > 3, the minimum detection limit assessed by direct on column injection was < 10 pg for analyte. The assays were linear from basal concentrations to 400 ng/ml. The intra- and inter-assay variations were < 10 and 15%, respectively. The assay has been applied successfully to measure plasma concentrations of these catecholamines in humans, rabbits and rats.
建立了一种反相高效液相色谱法结合电化学检测法,用于同时测定血浆中去甲肾上腺素、肾上腺素、多巴胺和3,4 -二羟基苯乙酸(DOPAC)的水平。通过由7%甲醇、0.1 M柠檬酸缓冲液(含0.3 mM乙二胺四乙酸钠(EDTA))和0.5 mM 1 -辛烷磺酸组成的流动相,在等度条件下以1.2 ml/min的流速运行,实现了儿茶酚胺和内标异丙肾上腺素的分离。保护池的电位设定为+650 mV,分析池的第一电极电位设定为+100 mV,第二电极电位设定为+350 mV。使用信噪比>3,通过直接柱上进样评估的最低检测限对于分析物小于10 pg。该测定法在基础浓度至400 ng/ml范围内呈线性。批内和批间变异分别<10%和15%。该测定法已成功应用于测定人、兔和大鼠血浆中这些儿茶酚胺的浓度。