Clinical evaluation and verification of the hyperthermia treatment planning system hyperplan.

作者信息

Gellermann J, Wust P, Stalling D, Seebass M, Nadobny J, Beck R, Hege H, Deuflhard P, Felix R

机构信息

Center of Radiation Medicine, Charite Medical School, Campus Virchow-Klinikum, Humboldt University at Berlin, Berlin, Germany.

出版信息

Int J Radiat Oncol Biol Phys. 2000 Jul 1;47(4):1145-56. doi: 10.1016/s0360-3016(00)00425-9.

Abstract

PURPOSE

A prototype of the hyperthermia treatment planning system (HTPS) HyperPlan for the SIGMA-60 applicator (BSD Medical Corp., Salt Lake City, Utah, USA) has been evaluated with respect to clinical practicability and correctness.

MATERIALS AND METHODS

HyperPlan modules extract tissue boundaries from computed tomography (CT) images to generate regular and tetrahedral grids as patient models, to calculate electric field (E-field) distributions, and to visualize three-dimensional data sets. The finite difference time-domain (FDTD) method is applied to calculate the specific absorption rate (SAR) inside the patient. Temperature distributions are calculated by a finite-element code and can be optimized. HyperPlan was tested on 6 patients with pelvic tumors. For verification, measured SAR values were compared with calculated SAR values. Furthermore, intracorporeal E-field scans were performed and compared with calculated profiles.

RESULTS

The HTPS can be applied under clinical conditions. Measured absolute SAR (in W/kg), as well as relative E-field scans, correlated well with calculated values (+/-20%) using the contour-based FDTD method. Values calculated by applying the FDTD method directly on the voxel (CT) grid, were less well correlated with measured data.

CONCLUSION

The HyperPlan system proved to be clinically feasible, and the results were quantitatively and qualitatively verified for the contour-based FDTD method.