Radiation dosimetry using polymer gels: methods and applications.

作者信息

McJury M, Oldham M, Cosgrove V P, Murphy P S, Doran S, Leach M O, Webb S

机构信息

Joint Department of Physics, Institute of Cancer Research, Sutton, Surrey, UK.

出版信息

Br J Radiol. 2000 Sep;73(873):919-29. doi: 10.1259/bjr.73.873.11064643.

Abstract

New, complex radiotherapy delivery techniques require dosimeters that are able to measure complex three-dimensional dose distributions accurately and with good spatial resolution. Polymer gel is an emerging new dosimeter being applied to these challenges. The aim of this review is to present a practical overview of polymer gel dosimetry, including gel manufacture, imaging, calibration and application to radiotherapy verification. The dosimeters consist of a gel matrix within which is suspended a solution of acrylic molecules. These molecules polymerize upon exposure to radiation, with the degree of polymerization being proportional to absorbed dose. The polymer distribution can be measured in two or three dimensions using MRI or optical tomography and, after calibration, the images can be converted into radiation dose distributions. Manufacture of the gel is reported to be reproducible, and measured dose in the range 0-10 Gy is accurate to within 3-5%. In-plane image resolution of 1 mm x 1 mm, with image slice thicknesses of between 2-5 mm, is typically achievable using clinical 1.5 T MR scanners and standard T2 weighted imaging sequences. The gels have been used to verify a number of conventional and novel radiotherapy modalities, including brachytherapy, intensity modulated radiotherapy and stereotactic radiosurgery. All the studies have confirmed the value and versatility of the dosimetry technique.

摘要

新的、复杂的放射治疗技术需要能够精确测量复杂三维剂量分布并具有良好空间分辨率的剂量仪。聚合物凝胶是一种正在应用于这些挑战的新兴剂量仪。本综述的目的是对聚合物凝胶剂量测定法进行实际概述,包括凝胶制造、成像、校准以及在放射治疗验证中的应用。剂量仪由凝胶基质组成,其中悬浮着丙烯酸分子溶液。这些分子在受到辐射时会聚合,聚合程度与吸收剂量成正比。聚合物分布可以使用磁共振成像(MRI)或光学断层扫描在二维或三维中进行测量,并且在校准后,图像可以转换为辐射剂量分布。据报道,凝胶的制造具有可重复性,在0 - 10 Gy范围内测量的剂量精确到3 - 5%以内。使用临床1.5 T磁共振扫描仪和标准T2加权成像序列通常可以实现1 mm×1 mm的平面图像分辨率,图像切片厚度在2 - 5 mm之间。这些凝胶已被用于验证多种传统和新型放射治疗方式,包括近距离放射治疗、调强放射治疗和立体定向放射外科手术。所有研究都证实了该剂量测定技术的价值和多功能性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验