King R, Bonfiglio R, Fernandez-Metzler C, Miller-Stein C, Olah T
Department of Drug Metabolism, West Point, Pennsylvania 19468, USA.
J Am Soc Mass Spectrom. 2000 Nov;11(11):942-50. doi: 10.1016/S1044-0305(00)00163-X.
We show results from experiments designed to determine the relative importance of gas phase processes and solution phase processes into ionization suppression observed in biological sample extracts. The data indicate that gas phase reactions leading to the loss of net charge on the analyte is not likely to be the most important process involved in ionization suppression. The results point to changes in the droplet solution properties caused by the presence of nonvolatile solutes as the main cause of ionization suppression in electrospray ionization of biological extracts.
我们展示了旨在确定气相过程和溶液相过程对生物样品提取物中观察到的电离抑制的相对重要性的实验结果。数据表明,导致分析物净电荷损失的气相反应不太可能是电离抑制中最重要的过程。结果表明,生物提取物电喷雾电离中电离抑制的主要原因是由非挥发性溶质的存在引起的液滴溶液性质的变化。