Blumenthal E Z, Williams J M, Weinreb R N, Girkin C A, Berry C C, Zangwill L M
Glaucoma Center and Diagnostic Imaging Laboratory, Department of Ophthalmology, University of California, San Diego, USA.
Ophthalmology. 2000 Dec;107(12):2278-82. doi: 10.1016/s0161-6420(00)00341-9.
To evaluate the reproducibility of optical coherence tomograph (OCT) retinal nerve fiber layer (RNFL) measurements in normal and glaucomatous eyes by means of the commercially available OCT 2000 instrument (Humphrey Systems, Dublin, CA).
Prospective instrument validation study.
One eye each from 10 normal subjects and 10 glaucoma patients.
Twenty subjects underwent a total of eight scanning sessions during two independent visits. In each session, five circular scans centered on the optic nerve head were performed. The first two sessions were performed by two experienced technicians. Followed by a 30-minute break, a third and a fourth session was completed by the same technicians. This sequence was duplicated on a second visit. Intrasession, intersession, intervisit, and interoperator reproducibility of quadrant and global RNFL measurements were calculated by use of a components of variance model.
RNFL thickness.
The coefficient of variation for the mean RNFL thickness was significantly smaller (P = 0.02) in normal eyes (6.9%) than in glaucoma eyes (11.8%). The estimated root mean squared error based on the statistical model using three scans per patient was 5.8 and 8.0 micrometer for normal and glaucoma eyes, respectively. A components of variance model showed most of the variance (79%) to be due to differences between patients. Only a modest contribution to variability was found for session (1%), visit (5%), and operator (2%).
With the commercially available OCT, our results indicate that the RNFL measurements are reproducible for both normal and glaucomatous eyes.
通过使用市售的OCT 2000仪器(Humphrey Systems,加利福尼亚州都柏林)评估正常眼和青光眼眼中光学相干断层扫描(OCT)视网膜神经纤维层(RNFL)测量的可重复性。
前瞻性仪器验证研究。
10名正常受试者和10名青光眼患者各一只眼。
20名受试者在两次独立就诊期间共进行了8次扫描。每次扫描时,以视乳头为中心进行5次环形扫描。前两次扫描由两名经验丰富的技术人员完成。休息30分钟后,同一名技术人员完成第三次和第四次扫描。该流程在第二次就诊时重复。通过使用方差成分模型计算象限和整体RNFL测量的扫描内、扫描间、就诊间和操作者间的可重复性。
RNFL厚度。
正常眼平均RNFL厚度的变异系数(6.9%)显著小于青光眼眼(11.8%)(P = 0.02)。基于每位患者三次扫描的统计模型估计的均方根误差,正常眼和青光眼眼分别为5.8和8.0微米。方差成分模型显示,大部分变异(79%)是由于患者之间的差异。扫描(1%)、就诊(5%)和操作者(2%)对变异性的贡献较小。
使用市售的OCT,我们的结果表明,RNFL测量在正常眼和青光眼眼中都是可重复的。