Bascom Palmer Eye Institute, Miller School of Medicine, University of Miami, Miami, Florida 33136, USA.
Invest Ophthalmol Vis Sci. 2010 Nov;51(11):5724-30. doi: 10.1167/iovs.10-5222. Epub 2010 Jun 23.
To assess the reproducibility of peripapillary retinal nerve fiber layer (RNFL) thickness and optic nerve head (ONH) parameters measured with Cirrus HD-OCT in glaucomatous eyes.
Fifty-five glaucomatous eyes were included in the study. The optic disc cube 200 × 200 protocol was used to obtain three scans during the same visit to evaluate the intravisit reproducibility. One scan on 4 additional days within a 2-month period of the first session was obtained to assess intervisit reproducibility. Intraclass correlation coefficient (ICC), coefficient of variation (CV), and test-retest SD (TRT SD) were calculated for each RNFL and ONH parameter. The formula 1.645 × √2 × intervisit TRT SD provides an upper tolerance limit to variability beyond which nonphysiologic change should be considered.
All ICCs were excellent, ranging from 83.9% to 99.2% for intravisit measurements and from 80.8% to 99.1% for intervisit measurements. Cup/disc area ratio had the lowest CV (1.1%) in either type of measurement, followed by average RNFL thickness (1.9% and 2.7%). Nasal clock hours and quadrants showed the poorest reproducibility as did the clock hour directly temporally. The intervisit tolerance limit for average RNFL thickness was 3.89 μm.
Intravisit and intervisit measurements of peripapillary RNFL thickness and ONH parameters with Cirrus HD-OCT showed excellent reproducibility, indicating that this instrument may be useful in monitoring glaucoma progression. When comparing two measurements from the same eye on two different visits, a reproducible decrease in average RNFL thickness of approximately 4 μm or more may be considered a statistically significant change from baseline.
评估 Cirrus HD-OCT 测量青光眼患者视盘周围视网膜神经纤维层(RNFL)厚度和视盘(ONH)参数的可重复性。
本研究纳入 55 只青光眼眼。采用视神经盘立方 200×200 方案,在同一次就诊时获得 3 次扫描,以评估日内可重复性。在第一次就诊后的 2 个月内的 4 天内获得 1 次额外扫描,以评估间日可重复性。计算每个 RNFL 和 ONH 参数的组内相关系数(ICC)、变异系数(CV)和测试-再测试标准差(TRT SD)。公式 1.645×√2×间日 TRT SD 提供了一个可接受的上限,超过这个上限应考虑非生理性变化。
所有 ICC 均非常优秀,日内测量的 ICC 范围为 83.9%至 99.2%,间日测量的 ICC 范围为 80.8%至 99.1%。在任何一种测量中,杯/盘面积比的 CV 最低(1.1%),其次是平均 RNFL 厚度(1.9%和 2.7%)。鼻时钟小时和象限的可重复性最差,直接颞侧时钟小时的可重复性也最差。平均 RNFL 厚度的间日可接受上限为 3.89μm。
Cirrus HD-OCT 测量视盘周围 RNFL 厚度和 ONH 参数的日内和间日测量具有极好的可重复性,表明该仪器可能有助于监测青光眼进展。当比较两次不同就诊时同一眼的两次测量时,平均 RNFL 厚度可接受的、大约 4μm 或更多的可重复性下降可被认为是与基线相比具有统计学意义的变化。