Reproducibility of spectral-domain optical coherence tomography measurements in adult and pediatric glaucoma.

作者信息

Ghasia Fatema F, El-Dairi Mays, Freedman Sharon F, Rajani Anand, Asrani Sanjay

机构信息

Duke Eye Center, Durham, NC.

出版信息

J Glaucoma. 2015 Jan;24(1):55-63. doi: 10.1097/IJG.0b013e31829521db.

Abstract

PURPOSE

To determine the reproducibility of spectral-domain optical coherence tomography (OCT) measurements of peripapillary retinal nerve fiber layer (RNFL) and macular retinal thickness of the central 20-degree area of the posterior pole in adult and pediatric glaucoma and normal subjects.

STUDY DESIGN

Prospective.

METHODS

Normal, glaucoma suspect and glaucoma eyes of adult and pediatric subjects were recruited. Spectralis OCT scans were obtained 3 times on the same day (intravisit) and 4 weeks later (intervisit). Six individual sectors, average RNFL, and macular retinal thickness using an 8×8 mm grid centered on the foveal pit were measured for all the eyes. The central 16 square areas centered on the foveal pit were called the central macular area and the surrounding 48 square areas were called the perimacular area. Intraclass correlation coefficient, coefficient of variation (COV), and test-retest variability were calculated.

RESULTS

Included were 51 adult eyes and 25 pediatric eyes. The intravisit and intervisit intraclass correlation coefficient for the central macular, perimacula, and average RNFL was very good (>0.8) for all adult and pediatric normal and glaucomatous eyes. The mean COVs of normal and glaucomatous eyes were higher for average RNFL (1.5 ± 1.5% adults, 1.0 ± 0.9% pediatric) compared to the central macula (0.56 ± 0.54% adults, 0.68 ± 0.64% pediatric) and perimacular thickness (0.64 ± 0.51% adults, 0.57 ± 0.42% pediatric) (1-way analysis of variance, P<0.05). No difference was noted in the intravisit and intervisit COVs between the adult and pediatric eyes of central macula, perimacula, and average RNFL thickness (unpaired t test, P>0.05). No intervisit reproducibility analysis was available for pediatric glaucoma eyes.

CONCLUSION

The reproducibility of Spectralis spectral-domain-OCT were very good for both RNFL and macular thickness measurements in glaucoma/glaucoma suspect eyes of adult and pediatric subjects.

摘要

目的

确定在成人和儿童青光眼患者及正常受试者中,使用光谱域光学相干断层扫描(OCT)测量视乳头周围视网膜神经纤维层(RNFL)以及后极部中心20度区域黄斑视网膜厚度的可重复性。

研究设计

前瞻性研究。

方法

招募成人和儿童受试者的正常眼、青光眼可疑眼和青光眼患眼。在同一天(就诊内)和4周后(就诊间)进行3次Spectralis OCT扫描。对所有眼睛测量六个独立扇形区域、平均RNFL以及以黄斑中心凹为中心的8×8 mm网格的黄斑视网膜厚度。以黄斑中心凹为中心的16个正方形区域称为中心黄斑区,周围48个正方形区域称为黄斑周边区。计算组内相关系数、变异系数(COV)和重测变异性。

结果

纳入51只成人眼和25只儿童眼。所有成人和儿童正常眼及青光眼患眼中,中心黄斑、黄斑周边和平均RNFL的就诊内和就诊间组内相关系数均非常好(>0.8)。与中心黄斑(成人0.56±0.54%,儿童0.68±0.64%)和黄斑周边厚度(成人0.64±0.51%,儿童0.57±0.42%)相比(单因素方差分析,P<0.05),正常眼和青光眼患眼中平均RNFL的平均COV更高(成人1.5±1.5%,儿童1.0±0.9%)。中心黄斑、黄斑周边和平均RNFL厚度的成人眼和儿童眼之间,就诊内和就诊间COV无差异(独立样本t检验,P>0.05)。儿童青光眼患眼无法进行就诊间可重复性分析。

结论

Spectralis光谱域OCT在成人和儿童青光眼/青光眼可疑患眼中测量RNFL和黄斑厚度的可重复性非常好。