The purpose of this study was to investigate the possibility that some artifactual high signals produced in CSF with fluid-attenuated inversion-recovery MR sequences could be due to inhomogeneity in the amplitude of the initial inversion pulse, and that this problem could be reduced or eliminated by the use of adiabatic inversion pulses. Studies with four volunteers showed dependence of high CSF signals in the posterior fossa on radiofrequency pulse amplitudes and that these signals could be eliminated by the use of adiabatic inversion pulses. Two illustrative clinical cases are included.
本研究的目的是调查脑脊液在液体衰减反转恢复磁共振序列中产生的一些伪高信号可能是由于初始反转脉冲幅度不均匀所致的可能性,以及使用绝热反转脉冲可以减少或消除这个问题。对四名志愿者的研究表明,后颅窝中脑脊液高信号与射频脉冲幅度有关,并且使用绝热反转脉冲可以消除这些信号。文中包含两个说明性临床病例。