Potential complication of bioabsorbable screw fixation for osteochondritis dissecans of the knee.

作者信息

Scioscia T N, Giffin J R, Allen C R, Harner C D

机构信息

Department of Orthopaedics, the University of Pittsburgh, Pittsburgh, Pennsylvania, U.S.A.

出版信息

Arthroscopy. 2001 Feb;17(2):E7. doi: 10.1053/jars.2001.17995.

Abstract

The 3 cases presented describe loosening or failure of bioabsorbable screws in the treatment of osteochondritis dissecans (OCD). In case 1, a 17-year-old boy with OCD of the medial femoral condyle was treated with bioabsorbable screw fixation. Six months after surgery, the patient had an acute episode of pain with effusion. Arthroscopic examination revealed 2 of the 4 screws backed out, causing cartilage damage to the corresponding tibial plateau. The shafts of the remaining 2 screws had completely absorbed, leaving the unabsorbed screw heads as intra-articular loose bodies in the knee. Unpredictable and inconsistent degradation of the screws is believed to be the mechanism for screw back-out and cartilage damage.

摘要

所呈现的3例病例描述了可吸收螺钉在治疗剥脱性骨软骨炎(OCD)时出现松动或失效的情况。病例1中,一名17岁患有股骨内侧髁剥脱性骨软骨炎的男孩接受了可吸收螺钉固定治疗。术后6个月,患者出现急性疼痛伴积液。关节镜检查发现4枚螺钉中有2枚退出,导致相应胫骨平台软骨损伤。其余2枚螺钉的杆部已完全吸收,未吸收的螺钉头作为膝关节内的游离体留存。螺钉不可预测且不一致的降解被认为是螺钉退出和软骨损伤的机制。