Clin Biochem. 1975 Apr;8(2):118-23. doi: 10.1016/s0009-9120(75)91206-0.
The inclusion of a minimum of 2 reference samples in the radioimmunoassay of testesterone and digoxin, and the competitive protein binding assay of cortisol served as a simple and practical index of overall assay performance. Statistical parameters, the mean plus or minus 2 standard deviation limits as well as the % coefficient of variation were determined at 3 clinically significant values for these samples. While a 50% intercept served mainly to indicate standard curve sensitivity, quality control sera provided a valid assessment of annay precision.