Liu L, Hu J, Wang L, Liu L, Zhou X
Department of Analysis-Measurement Science, Wuhan University, PR of China.
Fresenius J Anal Chem. 2001 Mar;369(6):483-5. doi: 10.1007/s002160000657.
A novel piezoelectric immunosensor has been developed for the detection of human complement C1-inhibitor. Anti-C1-inhibitor antibody was immobilized onto the gold electrodes of a 9 MHz AT-cut piezoelectric crystal. Coating the crystal with polyethyleneimine adhesion, followed by a glutaraldehyde cross-linking method to immobilize antibody showed better results than the physical adsorption method with respect to sensitivity and reproducibility. Under the optimized experimental conditions, the sensor showed good response to the C1-inhibitor in the range from 2.0 x 10(-8) to 1.2 x 10(-6) g. Other proteins in human serum did not remarkably interfere with the detection. The crystals could be regenerated 5 times, when bound materials on the crystal surface were eluted by strong acid and strong alkali solution and subsequently cleaned in an ultrasonic cleaner.
一种用于检测人补体C1抑制因子的新型压电免疫传感器已被开发出来。抗C1抑制因子抗体被固定在一个9 MHz AT切型压电晶体的金电极上。相对于物理吸附法,用聚乙烯亚胺进行粘附涂层,随后采用戊二醛交联法固定抗体,在灵敏度和重现性方面显示出更好的结果。在优化的实验条件下,该传感器对C1抑制因子在2.0×10(-8)至1.2×10(-6) g范围内表现出良好的响应。人血清中的其他蛋白质对检测没有明显干扰。当晶体表面的结合物用强酸和强碱溶液洗脱,随后在超声波清洗器中清洗时,晶体可以再生5次。