Chu X, Lin Z H, Shen G L, Yu R Q
Institute of Chemometrics and Chemical Sensing Technology, Hunan University, Changsha, China.
Analyst. 1995 Dec;120(12):2829-32. doi: 10.1039/an9952002829.
A piezoelectric immunosensor has been developed for the determination of human IgM. The crystals are AT-cut and have a basic resonant frequency of 9 MHz. Immobilization of goat antihuman IgM antibodies to the crystals' surfaces was accomplished via a CNBr-activated copolymer coating of 2-hydroxyethyl methacrylate and methylmethacrylate. The IgM piezoelectric immunosensor can be used for the human IgM determination in the range 5-93 micrograms ml-1. The analytical results given by this approach were in satisfactory agreement with those given by the single radical immunodiffusion procedure. The sensitivity, specificity and reproducibility of this immunosensor were investigated. Further, the valent value of goat antihuman IgM antibody binding with human IgM antigen and the affinity constant of immunoreaction in this experimental system were studied. After washing with tetrahydrofuran, a crystal can be re-used 20 times without detectable loss of sensitivity.
已开发出一种用于测定人免疫球蛋白M(IgM)的压电免疫传感器。晶体为AT切割型,基本共振频率为9兆赫。通过用溴化氰活化的甲基丙烯酸2-羟乙酯和甲基丙烯酸甲酯共聚物涂层,将山羊抗人IgM抗体固定在晶体表面。该IgM压电免疫传感器可用于测定浓度范围为5 - 93微克/毫升的人IgM。该方法给出的分析结果与单向免疫扩散法给出的结果吻合良好。研究了这种免疫传感器的灵敏度、特异性和重现性。此外,还研究了该实验系统中山羊抗人IgM抗体与人类IgM抗原结合的价数以及免疫反应的亲和常数。用四氢呋喃洗涤后,一块晶体可重复使用20次,且灵敏度无明显损失。