Compared vulnerabilities to small cardiac motions between different cameras used for myocardial perfusion imaging.

作者信息

Salvadori Julien, Petegnief Yolande, Sabbah Remi, Morel Olivier, Boulahdour Hatem, Karcher Gilles, Marie Pierre-Yves, Imbert Laetitia

机构信息

Institut de Cancérologie de Lorraine, Université de Lorraine, 54000, Nancy, France.

CHU-Besançon, Université de Franche-Comté, Service de Médecine Nucléaire, 25000, Besançon, France.

出版信息

J Nucl Cardiol. 2019 Aug;26(4):1313-1322. doi: 10.1007/s12350-017-1175-6. Epub 2018 Jan 11.

Abstract

This phantom-based study was aimed to determine whether cardiac CZT-cameras, which provide an enhanced spatial resolution and image contrast compared to Anger cameras, are similarly affected by small cardiac motions. Translations of a left ventricular (LV) insert at half-SPECT acquisitions through six possible orthogonal directions and with 5- or 10-mm amplitude were simulated on the Discovery NM-530c and DSPECT CZT-cameras and on an Anger Symbia T2 camera equipped with an astigmatic (IQ.SPECT) or conventional parallel-hole collimator (Conv.SPECT). SPECT images were initially reconstructed as currently recommended for clinical routine. The heterogeneity in recorded activity from the 17 LV segments gradually increased between baseline and motions simulated at 5- and 10-mm amplitudes with all cameras, although being higher for Anger- than CZT-cameras at each step and resulting in a higher mean number of artifactual abnormal segments (at 10-mm amplitude, Conv.SPECT: 3.7; IQ.SPECT: 1.8, Discovery: 0.7, DSPECT: 0). However, this vulnerability to motion was markedly (1) decreased for Conv.SPECT reconstructed without the recommended Resolution Recovery algorithm and (2) increased for DSPECT reconstructed without the recommended cardiac model. CZT-cameras and especially the DSPECT appear less vulnerable to small cardiac motions than Anger-cameras although these differences are strongly dependent on reconstruction parameters.

摘要

这项基于体模的研究旨在确定与安杰相机相比,具有更高空间分辨率和图像对比度的心脏CZT相机是否同样受到心脏小幅度运动的影响。在Discovery NM-530c和DSPECT CZT相机以及配备散光(IQ.SPECT)或传统平行孔准直器(Conv.SPECT)的安杰Symbia T2相机上,模拟了左心室(LV)插入物在半SPECT采集中通过六个可能的正交方向且幅度为5毫米或10毫米的平移。SPECT图像最初按照目前临床常规推荐的方法进行重建。在所有相机上,从17个LV节段记录的放射性活度的异质性在基线和模拟的5毫米及10毫米幅度运动之间逐渐增加,尽管在每个步骤中安杰相机的异质性高于CZT相机,并且导致伪像异常节段的平均数量更高(在10毫米幅度时,Conv.SPECT:3.7;IQ.SPECT:1.8,Discovery:0.7,DSPECT:0)。然而,这种对运动的易感性在(1)未使用推荐的分辨率恢复算法重建的Conv.SPECT中显著降低,(2)未使用推荐的心脏模型重建的DSPECT中增加。CZT相机,尤其是DSPECT相机,似乎比安杰相机对心脏小幅度运动的敏感性更低,尽管这些差异强烈依赖于重建参数。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验