Attrill D C, Ashley P F
Unit of Prosthodontics, University of Manchester.
Br Dent J. 2001 Apr 28;190(8):440-3. doi: 10.1038/sj.bdj.4800998.
To compare the accuracy and repeatability of three diagnostic systems (DIAGNOdent, visual and radiographic) for occlusal caries diagnosis in primary molars.
Two examiner, in vitro, blinded study. Histological gold standard.
58 occlusal surfaces of primary molars examined in turn by two examiners using each of three diagnostic systems (DIAGNOdent, visual and radiographic). These results were compared with a histological gold standard. Sensitivity and specificity were calculated for each diagnostic system at a range of thresholds. Inter- and intra- examiner repeatability were calculated for each diagnostic system using the kappa statistic.
Sensitivity, specificity, inter and intra examiner repeatability for each diagnostic system.
The highest sensitivity values were provided by DIAGNOdent (0.77 and 0.80, examiners 1 and 2 respectively) however this was offset by a lower specificity (0.82 and 0.85) than all other systems with the exception of examiner 1 at V1 visual threshold. The DIAGNOdent gave the highest values of kappa for intra- and inter-examiner repeatability with the exception of intra-examiner repeatability for examiner 2 where visual diagnosis had the highest value of kappa.
The DIAGNOdent was the most accurate system tested for the detection of occlusal dentine caries in primary molars. The performance of the DIAGNOdent systems was not statistically significantly better than that achieved using visual examination for non-cavitated teeth (V1 and V2 threshold). DIAGNOdent may prove useful as a predictive clinical tool, however with appropriate training visual examination may offer similar results without the need for additional equipment.
比较三种诊断系统(DIAGNOdent、视觉检查和X线检查)对乳磨牙咬合面龋诊断的准确性和可重复性。
两名检查者进行的体外双盲研究。组织学为金标准。
两名检查者依次使用三种诊断系统(DIAGNOdent、视觉检查和X线检查)对58个乳磨牙咬合面进行检查。将这些结果与组织学金标准进行比较。计算每个诊断系统在一系列阈值下的敏感性和特异性。使用kappa统计量计算每个诊断系统的检查者间和检查者内可重复性。
每个诊断系统的敏感性、特异性、检查者间和检查者内可重复性。
DIAGNOdent的敏感性最高(检查者1和检查者2分别为0.77和0.80),然而,除检查者1在V1视觉阈值时外,其特异性(0.82和0.85)低于所有其他系统。除检查者2的检查者内可重复性外,DIAGNOdent在检查者内和检查者间可重复性方面的kappa值最高,检查者2的视觉诊断kappa值最高。
DIAGNOdent是检测乳磨牙咬合面牙本质龋的最准确系统。对于非龋洞型牙齿(V1和V2阈值),DIAGNOdent系统的性能在统计学上并不显著优于视觉检查。DIAGNOdent可能被证明是一种有用的预测性临床工具,然而,经过适当培训,视觉检查可能无需额外设备就能提供类似结果。