We have assessed the accuracy and precision of our flowing afterglow mass spectrometric method (FA-MS) for absolute measurements of deuterium abundance in water using standardised tap water/D(2)O mixtures within the D/H range from 155 ppm (local tap water) to 1104 ppm, prepared by the gravimetric method. The results of this study show that a precision and accuracy of 1% can be achieved for the deuterium abundance in water samples. This is quite adequate for the main application, which is the rapid, non-invasive measurement of total body water by deuterium analysis of breath water vapour using the D(2)O dilution method.
我们使用重量法制备了D/H范围从155 ppm(当地自来水)到1104 ppm的标准化自来水/D₂O混合物,评估了我们的流动余辉质谱法(FA-MS)用于绝对测量水中氘丰度的准确性和精密度。本研究结果表明,对于水样中的氘丰度,可实现1%的精密度和准确度。这对于主要应用来说已经足够,该主要应用是通过使用D₂O稀释法对呼出的水蒸气进行氘分析来快速、无创地测量全身水含量。