This study was undertaken to assess the comparability of body water compartment estimates in healthy volunteers using single and dual frequency bioelectrical impedance analysis (BIA) with established reference methods of tritium and NaBr dilution.
Total body water (TBW) was estimated in 10 healthy volunteers using single frequency (50 kHz) BIA (Bodystat 1500), dual frequency (5 and 200 kHz) BIA (Bodystat Dualscan 2005) and tritium dilution. Extracellular water (ECW) was measured with dual frequency BIA and NaBr dilution. BIA was performed using distal tetrapolar electrodes in the supine position. Venous blood was sampled for measurement of background concentrations of tritium and NaBr using a beta counter and high performance liquid chromatography respectively. 10 ml tritiated water (3.7 MBq) and 50 ml 5% NaBr solution were then injected intravenously and blood samples taken from the opposite arm every 45 min for 4.5 h for estimation of concentrations of tritium and NaBr.
There was good correlation (r(2)=0.76) between estimates of ECW using dual frequency BIA and NaBr dilution, with the former overestimating ECW by approximately 1 L. However, this difference varied systematically with body weight. Although TBW measurements obtained by single and dual frequency BIA correlated well with estimates using tritium dilution (r(2)=0.96 and 0.95 respectively), single frequency BIA underestimated TBW by approximately 1 L and dual frequency BIA by approximately 5 L compared to tritium dilution.
TBW measurements obtained using the single frequency BIA device were more accurate than those obtained using the dual frequency BIA device. Dual frequency BIA provided a reasonably accurate estimate of ECW.
本研究旨在通过单频和双频生物电阻抗分析(BIA),并与已确立的氚和溴化钠稀释参考方法进行比较,评估健康志愿者体内水分隔室估计值的可比性。
使用单频(50kHz)BIA(Bodystat 1500)、双频(5和200kHz)BIA(Bodystat Dualscan 2005)和氚稀释法,对10名健康志愿者的总体水(TBW)进行估计。使用双频BIA和溴化钠稀释法测量细胞外液(ECW)。BIA采用仰卧位的远端四极电极进行。分别使用β计数器和高效液相色谱法采集静脉血,测量氚和溴化钠的背景浓度。然后静脉注射10ml氚化水(3.7MBq)和50ml 5%溴化钠溶液,并在4.5小时内每隔45分钟从对侧手臂采集血样,以估计氚和溴化钠的浓度。
双频BIA估计的ECW与溴化钠稀释法之间具有良好的相关性(r² = 0.76),前者使ECW高估了约1L。然而,这种差异随体重呈系统性变化。虽然单频和双频BIA获得的TBW测量值与氚稀释法的估计值相关性良好(分别为r² = 0.96和0.95),但与氚稀释法相比,单频BIA使TBW低估了约1L,双频BIA使TBW低估了约5L。
使用单频BIA设备获得的TBW测量值比使用双频BIA设备获得的测量值更准确。双频BIA对ECW提供了合理准确的估计。