Hau J, Stadler R, Jenny T A, Fay L B
Nestlé Research Center, Nestlé Ltd., Vers-chez-les-Blanc, CH-1000 Lausanne 26, Switzerland.
Rapid Commun Mass Spectrom. 2001;15(19):1840-8. doi: 10.1002/rcm.444.
The interpretation of mass spectra is a key process during compound identification, and the combination of tandem mass spectrometry (MS/MS) with high-accuracy mass measurements may deliver crucial information on the identity of a compound. Obtaining accurate mass data of fragment ions in MS/MS reveals the particular problem of mass calibration when a lockmass, which is frequently used to obtain accurate masses in MS, is absent. An alternative technique is to recalibrate the MS/MS spectrum using a reference MS/MS spectrum acquired under the same conditions. We have tested and validated this approach using a hybrid quadrupole/orthogonal acceleration reflectron-type time-of-flight (TOF) mass spectrometer. The results were compared with those obtained under similar conditions on a Fourier transform ion cyclotron resonance (FT-ICR) instrument. We found that the mass accuracy observed with such an "external" recalibration on the TOF instrument in MS/MS is identical to what can be obtained on a similar instrument operating in one-dimensional MS mode using the lockmass technique. However, mass accuracy in both cases is one order of magnitude inferior to that obtained using FTMS, and also inferior to that observed using sector field MS when operated at comparable resolution. Nevertheless, for small (<200 Da) molecules, this mass accuracy was still sufficient to have the "true" elemental composition identified as the first hit in about 70% of all cases. It was possible to elucidate the fragmentation mechanism of eight azaheterocycles containing a pyridine moiety, where the accurate mass data from the TOF instrument allowed distinction between two alternative fragmentation pathways.
质谱的解析是化合物鉴定过程中的关键步骤,串联质谱(MS/MS)与高精度质量测量相结合可提供有关化合物身份的关键信息。当缺少常用于在MS中获取准确质量的锁定质量时,在MS/MS中获取碎片离子的准确质量数据会揭示质量校准的特殊问题。一种替代技术是使用在相同条件下采集的参考MS/MS谱图对MS/MS谱图进行重新校准。我们使用混合四极杆/正交加速反射式飞行时间(TOF)质谱仪对该方法进行了测试和验证。将结果与在傅里叶变换离子回旋共振(FT-ICR)仪器上在类似条件下获得的结果进行了比较。我们发现,在TOF仪器上MS/MS中通过这种“外部”重新校准观察到的质量精度与在使用锁定质量技术的一维MS模式下运行的类似仪器上可获得的质量精度相同。然而,这两种情况下的质量精度都比使用FTMS获得的精度低一个数量级,并且也低于在可比分辨率下运行的扇形场MS所观察到的精度。尽管如此,对于小分子(<200 Da),这种质量精度在约70%的所有情况下仍足以将“真实”元素组成作为首次命中结果识别出来。有可能阐明八个含有吡啶部分的氮杂环的碎裂机制,其中TOF仪器的准确质量数据允许区分两种替代的碎裂途径。