Measurement of systolic blood pressure in fingers of newborn infants.

作者信息

Gundersen J, Dahlin K

出版信息

Acta Paediatr Scand. 1975 Sep;64(5):441-4. doi: 10.1111/j.1651-2227.1975.tb03913.x.

Abstract

The systolic blood pressure in the index finger was measured using a minicuff and a strain gauge both placed around the finger. 122 measurements in four newborn infants were performed during monitoring in an incubator and the systolic blood pressure values obtained were compared with the simultaneously recorded intra-arterial blood pressure in the aorta. Good correlation between direct and indirect pressure values was found. An important advantage is that the measurements do not arouse the infants nor disturb their sleep. The procedure seems suited for further technical development and automatic measurements.

摘要

使用一个小型袖带和一个应变仪环绕手指来测量食指的收缩压。在保温箱监测期间,对4名新生儿进行了122次测量,并将获得的收缩压值与同时记录的主动脉内动脉血压进行比较。发现直接和间接压力值之间具有良好的相关性。一个重要的优点是这些测量不会唤醒婴儿,也不会干扰他们的睡眠。该方法似乎适合进一步的技术开发和自动测量。