Iyriboz Y
Department of Kinesiology, Louisiana State University, Baton Rouge 70803.
J Fam Pract. 1990 Oct;31(4):376-80.
In this study, a recently marketed proprietary finger blood pressure monitor, the Marshall, Astro F-88, was compared with the standard auscultative brachial mercury sphygmomanometer on 125 subjects. Measurements were undertaken according to the standards set by the American Heart Association. Sensitivity of the finger blood pressure measurement was 76% for systolic and 75% for diastolic blood pressure in diagnosis of high blood pressure (systolic greater than 140 mm Hg and diastolic greater than 90 mm Hg). Specificity was 86% for systolic and 82% for diastolic blood pressure. Positive predictive values were 58% for systolic and 38% for diastolic blood pressure in the study population in which prevalence of hypertension was 12%. The correlation coefficient (Pearson) for systolic values between devices was 0.76 (P less than .0001) and 0.57 (P less than .0001) for diastolic pressure. Values obtained by the finger monitor were found to be higher than those obtained by the mercury sphygmomanometer. Mean differences and standard deviations (paired t test) for systolic and diastolic pressures between the two devices were 2.3 +/- 14.9 mm Hg (P less than .08) and 2.9 +/- 14.5 mm Hg (P less than .02), respectively. These values are not in accordance with the proposed national standards because only 48% of the systolic and 37% of the diastolic blood pressure measurements were within 5 mm Hg of the mercury sphygmomanometer measurements. Therefore, although these differences may well be due to different techniques of monitoring employed by the devices, this device is not recommended for evaluation of blood pressure.
在本研究中,将一款近期上市的专利手指血压监测仪Marshall Astro F - 88与标准听诊式肱动脉水银血压计对125名受试者进行了比较。测量是根据美国心脏协会制定的标准进行的。在高血压诊断(收缩压大于140 mmHg且舒张压大于90 mmHg)中,手指血压测量的收缩压敏感性为76%,舒张压敏感性为75%。收缩压特异性为86%,舒张压特异性为82%。在高血压患病率为12%的研究人群中,收缩压的阳性预测值为58%,舒张压的阳性预测值为38%。两种设备之间收缩压值的相关系数(Pearson)为0.76(P小于0.0001),舒张压为0.57(P小于0.0001)。发现手指监测仪获得的值高于水银血压计获得的值。两种设备之间收缩压和舒张压的平均差异及标准差(配对t检验)分别为2.3±14.9 mmHg(P小于0.08)和2.9±14.5 mmHg(P小于0.02)。这些值不符合拟议的国家标准,因为只有48%的收缩压测量值和37%的舒张压测量值在水银血压计测量值的5 mmHg范围内。因此,尽管这些差异很可能是由于设备采用的不同监测技术所致,但不建议使用该设备评估血压。