在骨水泥型全髋关节置换术中,将三角形远端扶正器插入柄尖时的远端骨水泥套厚度。

Distal cement mantle thickness with a triangular distal centralizer inserted into the stem tip in cemented total hip arthroplasty.

作者信息

Kawate K, Ohmura T, Nakajima H, Takakura Y

机构信息

Department of Orthopaedic Surgery, Nara Medical University, Kashihara, Nara, Japan.

出版信息

J Arthroplasty. 2001 Dec;16(8):998-1003. doi: 10.1054/arth.2001.26592.

Abstract

Fifty-seven hips undergoing a primary cemented total hip arthroplasty with use of a triangular distal centralizer inserted into the stem tip were investigated, with a special focus on distal cement mantle thickness. The subjects were 43 women and 3 men. When evaluated on conventional anteroposterior and lateral radiographs, the relative incidence of thin cement mantles in zones 3 or 5 owing to the distal centralizer was 15.8% (9 hips). Unsatisfactory findings were that the distal centralizer and distal part of the stem shifted to the cortex in 7 hips and bent or failed at the inserted site in 2 hips. There is risk of generating a thin cement mantle with use of the triangular centralizer and its insertion into the stem tip.

摘要

对57例使用插入柄尖的三角形远端中心化器进行初次骨水泥型全髋关节置换术的髋关节进行了研究,特别关注远端骨水泥套厚度。受试者包括43名女性和3名男性。在常规前后位和侧位X线片上评估时,由于远端中心化器导致3区或5区骨水泥套变薄的相对发生率为15.8%(9髋)。令人不满意的发现是,7髋的远端中心化器和柄的远端部分向皮质移位,2髋在插入部位弯曲或失效。使用三角形中心化器并将其插入柄尖有产生薄骨水泥套的风险。