Continuous, simultaneous measurement of arterial oxygen tension and arterial blood pressure.

作者信息

Gold M I, Duarte I

出版信息

J Thorac Cardiovasc Surg. 1975 Oct;70(4):656-60.

PMID:1177478
Abstract

This is an investigation of a technique for simultaneously measuring arterial oxygen tension (PaO2) and blood pressure continuously during and after surgical anesthesia. In 27 patients, a special T attachment was placed on the end of an indwelling arterial catheter so that pressurized dilute heparin could be flushed through continuously while blood pressure was being recorded by means of a transducer-oscilloscope system and PaO2 was being recorded through an indwelling polarograph oxygen tension (PO2) readout device. The electrode was left in the artery for an average of 16 hours; the average PaO2 during mechanical ventilation with the patient breathing 50 per cent oxygen was approximately 200 torr at the beginning of anesthesia; it dropped to 158 torr 6 hours later. The average PaO2 in the recovery room 1 hour postoperatively, with the patient spontaneously breathing approximately 30 per cent oxygen, was 122 torr. No significant complications were encountered during this investigation. We believe this dual technique to be a valuable monitoring tandem in the anesthetized and surgical intensive care patient.

摘要

这是一项关于在手术麻醉期间及术后连续同步测量动脉血氧分压(PaO₂)和血压的技术的研究。对27例患者,在留置动脉导管末端放置一个特殊的T形接头,以便在通过换能器 - 示波器系统记录血压并通过留置极谱式氧分压(PO₂)读数装置记录PaO₂时,能持续冲洗加压稀释肝素。电极留在动脉内的平均时间为16小时;麻醉开始时,患者吸入50%氧气进行机械通气期间的平均PaO₂约为200托;6小时后降至158托。术后1小时在恢复室,患者自主呼吸约30%氧气时的平均PaO₂为122托。在这项研究中未遇到明显并发症。我们认为这种双重技术对于麻醉和外科重症监护患者是一种有价值的监测组合。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验