Alumina-on-alumina total hip arthroplasty: a minimum 18.5-year follow-up study.

作者信息

Hamadouche Moussa, Boutin Pierre, Daussange Jacques, Bolander Mark E, Sedel Laurent

机构信息

Laboratoire de Recherches Orthopédiques, Faculté de Médecine Lariboisière St. Louis, Université D. Diderot Paris VII, France.

出版信息

J Bone Joint Surg Am. 2002 Jan;84(1):69-77.

Abstract

BACKGROUND

The purpose of this retrospective study was to report the results, after a minimum of 18.5 years of follow-up, in a consecutive series of total hip arthroplasties performed with an alumina-on-alumina combination.

METHODS

One hundred and eighteen consecutive total hip arthroplasties were performed in 106 patients between 1979 and 1980. The prostheses combined a 32-mm alumina head and an all-alumina socket. Both components were cemented in eighty-five hips, both components were implanted without cement in twenty-nine, and only the stem was cemented in four. The mean age of the patients at the time of the index arthroplasty was 62.2 years (range, thirty-two to eighty-nine years).

RESULTS

At the 18.5 to 20.5-year follow-up evaluation, forty-five patients (fifty-one hips) were alive and had not had a revision, twenty-five patients (twenty-five hips) had undergone revision of one or both components, twenty-seven patients (thirty hips) had died, and nine patients (twelve hips) had been lost to follow-up. The mean Merle d'Aubigné hip score (and standard deviation) was 16.2 +/- 1.8 points at the latest follow-up evaluation. The rate of survival at twenty years, with revision for any reason as the end-point, was 85.6% for the cementless cups compared with 61.2% for the cemented cups and 84.9% for the cementless stems compared with 87.3% for the cemented stems. Wear of the prosthetic components was undetectable on plain radiographs. Periprosthetic cystic or scalloped lesions requiring the use of allograft bone during revision were present in three of the twenty-five revised hips. In addition, seven hips had moderate acetabular osteolysis treated with a 4-mm-larger cup. No fracture of the alumina socket or head was recorded. The mean acetabular wear rate in this series was <0.025 mm/yr.

CONCLUSION

With the alumina-on-alumina total hip arthroplasty, minimal wear rates and limited osteolysis can be expected up to twenty years after the operation, provided that sound acetabular component fixation is obtained.

摘要

背景

本回顾性研究的目的是报告在至少18.5年的随访后,一系列连续进行的氧化铝对氧化铝组合全髋关节置换术的结果。

方法

1979年至1980年间,对106例患者连续实施了118例全髋关节置换术。假体采用32毫米氧化铝股骨头和全氧化铝髋臼杯组合。85例髋关节的两个组件均用骨水泥固定,29例未用骨水泥植入两个组件,4例仅股骨柄用骨水泥固定。初次置换术时患者的平均年龄为62.2岁(范围32至89岁)。

结果

在18.5至20.5年的随访评估中,45例患者(51髋)存活且未进行翻修,25例患者(25髋)对一个或两个组件进行了翻修,27例患者(30髋)死亡,9例患者(12髋)失访。在最近一次随访评估时,Merle d'Aubigné髋关节平均评分(及标准差)为16.2±1.8分。以任何原因翻修为终点,无骨水泥髋臼杯20年生存率为85.6%,有骨水泥髋臼杯为61.2%;无骨水泥股骨柄20年生存率为84.9%,有骨水泥股骨柄为87.3%。在普通X线片上未检测到假体组件磨损。25例翻修髋关节中有3例存在假体周围囊性或扇形病变,翻修时需要使用同种异体骨。此外,7髋有中度髋臼骨溶解,采用大4毫米的髋臼杯进行治疗。未记录到氧化铝髋臼杯或股骨头骨折。本系列髋臼平均磨损率<0.025毫米/年。

结论

采用氧化铝对氧化铝全髋关节置换术,若髋臼组件固定良好,术后20年内可预期磨损率极低且骨溶解有限。