Boscia Francesco, La Tegola Maria Gabriella, Alessio Giovanni, Sborgia Carlo
Dipartimento di Oftalmologia ed Otorinolaringoiatria, Università di Bari, Italy.
J Cataract Refract Surg. 2002 Feb;28(2):253-8. doi: 10.1016/s0886-3350(01)01162-2.
To compare corneal thickness measurements with the Orbscan instrument (Orbtek, Inc.) and an ultrasonic pachymeter (Alcon Surgical) in corneas with haze.
Department of Ophthalmology and Otolaryngology, University of Bari, Bari, Italy.
Six eyes of 6 patients having phototherapeutic keratectomy for treatment of grade 2+ haze were assessed preoperatively and postoperatively by Orbscan optical pachymetry and ultrasonic pachymetry. Contralateral transparent untreated corneas of the same patients were used as controls. Two Orbscan and 5 ultrasound pachymetry measurements were performed at each visit and the mean values recorded. The accuracy and repeatability of the measurements were evaluated.
Preoperatively, the median corneal thickness in the 6 eyes with haze was 512.7 microm by ultrasonic pachymetry and 375.0 microm by optical pachymetry. The mean difference of 143.1 microm was statistically significant (Wilcoxon signed rank test, P= .0277). Postoperatively, the median pachymetry was 439.0 microm and 433.0 microm, respectively; the mean difference was 10.7 microm (P= .0277). In the control group, the median ultrasonic pachymetry was 530.9 microm and the median optical pachymetry, 529.0 microm; the mean difference was 4.7 microm (P= .1730). Precision was good in all groups; the standard deviation was less than 6.0 microm.
The Orbscan system showed a decreased accuracy in measuring corneal thickness when clinically significant haze was present. The repeatability of the measures was good, suggesting that the cause of error lies in the optical acquisition process.
比较使用Orbscan仪器(Orbtek公司)和超声角膜测厚仪(爱尔康手术公司)测量有 haze 的角膜的厚度。
意大利巴里大学眼耳鼻喉科。
对6例因治疗2+级 haze 而接受光治疗性角膜切削术的患者的6只眼在术前和术后进行Orbscan光学测厚和超声测厚评估。将同一患者对侧未治疗的透明角膜作为对照。每次就诊时进行2次Orbscan测量和5次超声测厚,并记录平均值。评估测量的准确性和可重复性。
术前,6只 haze 眼的角膜厚度中位数通过超声测厚为512.7微米,通过光学测厚为375.0微米。平均差值143.1微米具有统计学意义(Wilcoxon符号秩检验,P = 0.0277)。术后,测厚中位数分别为439.0微米和433.0微米;平均差值为10.7微米(P = 0.0277)。在对照组中,超声测厚中位数为530.9微米,光学测厚中位数为529.0微米;平均差值为4.7微米(P = 0.1730)。所有组的精密度良好;标准差小于6.0微米。
当存在具有临床意义的 haze 时,Orbscan系统在测量角膜厚度方面显示出准确性下降。测量的可重复性良好,表明误差原因在于光学采集过程。