Marsich M W, Bullimore M A
The Ohio State University, College of Optometry, Columbus 43210, USA.
Cornea. 2000 Nov;19(6):792-5. doi: 10.1097/00003226-200011000-00007.
To compare the repeatability of three measures of corneal thickness: Orbscan Slitscan pachymetry, ultrasound pachymetry, and optical pachymetry.
Twenty normal subjects were tested on three occasions. Two occurred on the same day and the third was on a different day at approximately the same time of day as one of the first two visits. Central corneal thickness of the right eye was measured with a Haag-Streit optical pachymeter, a Humphrey Model 855 ultrasound pachymeter, and the Orbscan system. Day-to-day and same-day repeatability was assessed by calculating the difference between the values from two visits and determining the mean difference, the SD, and the 95% limits of agreement (LoA) (LoA = mean +/- 1.96 SD).
Mean (+/- SD) central corneal thickness as measured by each instrument was as follows: 539 +/- 33 microm for optical pachymetry, 542 +/- 33 microm for ultrasound pachymetry, and 596 +/- 40 microm for Orbscan pachymetry. For day-to-day comparisons, optical pachymetry showed the poorest repeatability with 95% LoA of -61 to +32 microm. Ultrasound pachymetry showed better repeatability with 95% LoA of -22 to +24 microm. The Orbscan showed the best repeatability centrally with 95% LoA of -10 to +17 microm. Peripheral Orbscan pachymetry was less repeatable than that measured centrally but still more repeatable than central optical pachymetry. Similar results were found with same-day comparisons.
The Orbscan system is the most repeatable technique for measuring corneal thickness but shows a significant bias toward greater corneal thickness measures than both ultrasound and optical pachymetry.
比较三种角膜厚度测量方法的可重复性:Orbscan 裂隙扫描测厚法、超声测厚法和光学测厚法。
对 20 名正常受试者进行了三次测试。其中两次测试在同一天进行,第三次测试在与前两次测试中某一次大致相同的时间,但在不同的日期。使用 Haag-Streit 光学测厚仪、Humphrey Model 855 超声测厚仪和 Orbscan 系统测量右眼的中央角膜厚度。通过计算两次测量值之间的差异,并确定平均差异、标准差和 95%一致性界限(LoA)(LoA = 平均值 +/- 1.96 标准差)来评估每日和同日的可重复性。
每种仪器测量的平均(+/-标准差)中央角膜厚度如下:光学测厚法为 539 +/- 33 微米,超声测厚法为 542 +/- 33 微米,Orbscan 测厚法为 596 +/- 40 微米。在每日比较中,光学测厚法的可重复性最差,95% LoA 为 -61 至 +32 微米。超声测厚法的可重复性较好,95% LoA 为 -22 至 +24 微米。Orbscan 在中央区域的可重复性最佳,95% LoA 为 -10 至 +17 微米。Orbscan 周边角膜测厚的可重复性低于中央测量,但仍比中央光学测厚法更具可重复性。同日比较也发现了类似结果。
Orbscan 系统是测量角膜厚度最具可重复性的技术,但与超声和光学测厚法相比,它对角膜厚度测量值存在明显的偏大偏差。