Comparison of Fourier-Domain and Time-Domain Optical Coherence Tomography in the Measurement of Thinnest Corneal Thickness in Keratoconus.

作者信息

Wang Chunxiao, Xia Xueying, Tian Bishan, Zhou Shiyou

机构信息

State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangzhou, Guangdong 510060, China.

Guangdong Women and Children Hospital, Guangzhou, Guangdong 510000, China.

出版信息

J Ophthalmol. 2015;2015:402925. doi: 10.1155/2015/402925. Epub 2015 Jun 25.

Abstract

Objective. To compare fourier-domain optical coherence tomography (FD-OCT) and time-domain OCT (TD-OCT) in the determination of thinnest corneal thickness (TCT). Methods. This study included 55 keratoconus patients and 50 healthy volunteers. The RTVue-OCT (FD-OCT) and Visante-OCT (TD-OCT) were used for the measurement of the TCT. Three consecutive scans were performed. The comparison and agreement between the two modalities were analyzed by paired t-test, the Pearson correlation, intraclass correlation coefficient (ICC), and Bland-Altman plots. intraobserver repeatability was analyzed by the intraobserver within-subject standard deviation (S (w)), coefficient of variation (CV(w)), and ICC. Results. The TCT value of normal corneas was higher by RTVue-OCT (530.4 ± 19.7 μm) than Visante-OCT (521.5 ± 18.3 μm) (p < 0.001). For keratoconus eyes, the TCT was 425.0 ± 58.2 μm and 424.4 ± 55.7 μm (difference being 0.6 ± 10.2 μm, p = 0.604). Strong correlation (r = 0.938∼0.985) (ICC = 0.915-0.984) was observed between the two OCTs, and each OCT exhibited excellent repeatability in determining the TCT in all subjects (ICC = 0.984-0.994). Conclusions. The values of TCT obtained from RTVue-OCT and Visante-OCT were highly correlated; however, the two values were different. Both OCT instruments exhibited good intraobserver reliability. The existence of systematic differences suggested that the two instruments cannot be used interchangeably.

摘要

目的。比较傅里叶域光学相干断层扫描(FD-OCT)和时域光学相干断层扫描(TD-OCT)在测定最薄角膜厚度(TCT)方面的差异。方法。本研究纳入了55例圆锥角膜患者和50名健康志愿者。使用RTVue-OCT(FD-OCT)和Visante-OCT(TD-OCT)测量TCT。进行连续三次扫描。通过配对t检验、Pearson相关性分析、组内相关系数(ICC)和Bland-Altman图分析两种方法之间的比较和一致性。观察者内重复性通过观察者内受试者标准差(S(w))、变异系数(CV(w))和ICC进行分析。结果。正常角膜的TCT值通过RTVue-OCT(530.4±19.7μm)测量高于Visante-OCT(521.5±18.3μm)(p<0.001)。对于圆锥角膜眼,TCT分别为425.0±58.2μm和424.4±55.7μm(差异为0.6±10.2μm,p = 0.604)。两种OCT之间观察到强相关性(r = 0.938~0.985)(ICC = 0.915 - 0.984),并且每种OCT在所有受试者中测定TCT时均表现出极好的重复性(ICC = 0.984 - 0.994)。结论。从RTVue-OCT和Visante-OCT获得的TCT值高度相关;然而,这两个值不同。两种OCT仪器均表现出良好的观察者内可靠性。系统差异的存在表明这两种仪器不能互换使用。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/9700/4496494/775bd9e677fc/JOPH2015-402925.001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验