Buehl Wolf, Findl Oliver, Menapace Rupert, Georgopoulos Michael, Rainer Georg, Wirtitsch Matthias, Siegl Hannes, Pinz Axel
Department of Ophthalmology, University of Vienna, Vienna, Austria.
J Cataract Refract Surg. 2002 Feb;28(2):265-70. doi: 10.1016/s0886-3350(01)01228-7.
To determine the short-term reproducibility of standardized digital retroillumination images of regeneratory posterior capsule opacification (PCO) using the photographic setup at 1 institution.
Department of Ophthalmology, University of Vienna, Vienna, Austria.
In this prospective study, 60 retroillumination images of 30 eyes with varying degrees of PCO and different types of intraocular lenses were acquired with a standardized digital coaxial retroillumination system. Two images were taken per eye with a 1-minute interval between images. Ten other eyes were photographed in the same way but with a 5-day interval between the 2 images. All images were evaluated with a fully automated, objective PCO analysis software in which the PCO score was from 0 to 100. The 2 results (A, B) in each eye were compared, and the differences were calculated.
There was a high correlation between the A and B results (r = 0.99). The mean absolute difference was 3.7%. The repeatability coefficient was 8.8%.
Digital coaxial retroillumination photography provided quick acquisition of regeneratory PCO images. It provided excellent image quality and high reproducibility. The technique forms a good basis for automated quantification of PCO with new software systems.
使用某一机构的摄影装置确定标准化数字后照法图像对再生性后囊膜混浊(PCO)的短期可重复性。
奥地利维也纳大学眼科。
在这项前瞻性研究中,使用标准化数字同轴后照法系统采集了30只患有不同程度PCO和不同类型人工晶状体眼的60张后照法图像。每只眼睛拍摄两张图像,图像之间间隔1分钟。另外10只眼睛以相同方式拍摄,但两张图像之间间隔5天。所有图像均使用全自动客观PCO分析软件进行评估,其中PCO评分为0至100。比较每只眼睛的两个结果(A、B),并计算差异。
A和B结果之间存在高度相关性(r = 0.99)。平均绝对差异为3.7%。重复性系数为8.8%。
数字同轴后照法摄影可快速获取再生性PCO图像。它提供了出色的图像质量和高可重复性。该技术为使用新软件系统对PCO进行自动定量分析奠定了良好基础。