Taylor P N, Tromans A M, Harris K R, Swain I D
Department of Medical Physics and Biomedical Engineering, Salisbury District Hospital, Salisbury, Wiltshire SP2 8BJ, UK.
Spinal Cord. 2002 Jan;40(1):34-6. doi: 10.1038/sj.sc.3101250.
Evaluation of a FES (Functional Electrical Stimulation) device for the relief of postural hypertension and augmentation of cough in a C3 ventilator-dependent tetraplegic.
A single case study.
A supra regional spinal unit in the UK.
A dual channel stimulator was designed that allowed selection and initiation of two predetermined stimulation intensities using a chin controlled joystick. Two sets of 70 mm diameter electrodes were placed either side of the abdomen. Approximately 80 mA, 300 micros, 40 Hz was required for assisted cough while about 40 mA was required for maintenance of blood pressure. After eating, the lower level stimulus is self-administered every 3 to 5 min gradually increasing the time between groups of burst to once every hour after 90 min.
Following eating, a blood pressure of 60/45 mmHg was recorded. After five 1 s bursts of stimulation in quick repetition, this was increased to 133/92 mmHg. After 2 min blood pressure had fallen to 124/86 mmHg and to 93/66 after a further 4 min. The electrical stimulation was then repeated, returning the blood pressure to the previous higher level. Measurement of peak expiratory flow showed an increase from 275 l/min for an unassisted cough to 425 l/min when using the device.
The device is used every day. The user is now independent in coughing function and no longer requires suction or manual assistance. Maintenance of blood pressure has significantly improved his quality of life.
评估一种功能性电刺激(FES)设备对一名依赖C3通气的四肢瘫痪患者缓解姿势性高血压及增强咳嗽能力的效果。
单病例研究。
英国一个超地区性脊髓病治疗中心。
设计了一种双通道刺激器,可通过下巴控制的操纵杆选择并启动两种预定的刺激强度。两组直径70毫米的电极置于腹部两侧。辅助咳嗽大约需要80毫安、300微秒、40赫兹的刺激,而维持血压大约需要40毫安。进食后,每3至5分钟自行施加较低强度的刺激,90分钟后逐渐将每组刺激脉冲之间的时间间隔增加到每小时一次。
进食后,记录到血压为60/45毫米汞柱。在快速重复进行5次1秒的刺激脉冲后,血压升至133/92毫米汞柱。2分钟后血压降至124/86毫米汞柱,再过4分钟后降至93/66毫米汞柱。然后重复电刺激,血压恢复到之前的较高水平。呼气峰值流量测量显示,未使用该设备时自主咳嗽的流量为每分钟275升,使用该设备时增至每分钟425升。
该设备每天都在使用。使用者现在咳嗽功能独立,不再需要抽吸或人工辅助。血压的维持显著改善了他的生活质量。