根管长度的体内测定:使用Tri Auto ZX根尖定位手机的初步报告。

In vivo determination of root canal length: a preliminary report using the Tri Auto ZX apex-locating handpiece.

作者信息

Grimberg F, Banegas G, Chiacchio L, Zmener O

机构信息

Department of Adult Dental Care, Faculty of Odontology, University of Buenos Aires, Argentina.

出版信息

Int Endod J. 2002 Jul;35(7):590-3. doi: 10.1046/j.1365-2591.2002.00505.x.

Abstract

AIM

The aim of this study was to assess the clinical perfomance of a cordless handpiece with a built-in apex locator - the Tri Auto ZX - designed for root canal preparation with nickel-titanium rotary files.

METHODOLOGY

Twenty-five human maxillary incisor and canine teeth scheduled for extraction with mature apices were selected for the study. Informed written consent was obtained from each patient before treatment. After administration of local anaesthesia, the teeth were isolated and the pulp cavities accessed. The Tri Auto ZX along with a size 15 K-file was used in its electronic apex locating function based on the manufacturer's recommendations. A periapical radiograph with the file at the electronically determined constriction was taken, the file removed and the measurement registered as the electronic length (EL). To test the auto reverse function, a size 20 ProFile.04 taper NiTi rotary instrument was mounted in the handpiece. The point for the auto apical reverse function was preset on the panel at the 0.5 mm level. After the file was introduced into the canal and reached the predetermined level, the file automatically stopped and rotated in the opposite direction. A reference point was marked and this measurement was registered as the auto reverse length (ARL). All measurements were made twice by two different investigators. Teeth were then extracted and immersed in a 20% formalin solution for 48 h. After fixation, a size 15 file was inserted into the canal to measure the actual root canal length from the same reference point obtained with the Tri Auto ZX to the apical foramen, as seen in the stereo microscope. When the file tip was visible at the anatomical end of the canal it was withdrawn 0.5 mm and this measurement was registered as the actual length (AL). All measurements were expressed in mm and the measuring accuracy was set to 0.5 mm. The significance of the mean differences between EL and ARL and between EL and AL measurements at the 5% confidence level was evaluated.

RESULTS

EL measurements were coincident to ARL in all instances. EL and ARL were coincident to AL in 10 (40%) canals, in the remaining 15 canals (60%) the AL measurements were longer than EL and ARL (+0.5 mm) in 14 instances and shorter (-0.5 mm) in one case. Overall, the AL was longer than the EL or ARL, the mean difference being -0.23 mm +/- 0.32 (P < 0.05).

CONCLUSIONS

It was concluded that the Tri Auto ZX was useful and reliable. The Tri Auto ZX measurements protected against overpreparation.

摘要

目的

本研究旨在评估一款内置根尖定位仪的无绳手机(Tri Auto ZX)的临床性能,该手机专为使用镍钛旋转锉进行根管预备而设计。

方法

选取25颗计划拔除的具有成熟根尖的人类上颌切牙和尖牙用于本研究。治疗前获得每位患者的书面知情同意书。局部麻醉后,隔离牙齿并打开髓腔。按照制造商的建议,将Tri Auto ZX与一根15号K锉一起用于其电子根尖定位功能。在锉位于电子确定的狭窄处时拍摄根尖片,取出锉并将测量值记录为电子长度(EL)。为测试自动反转功能,将一根20号ProFile.04锥度镍钛旋转器械安装在手机上。自动根尖反转功能的点在面板上预设为0.5毫米水平。将锉插入根管并到达预定水平后,锉自动停止并反向旋转。标记一个参考点并将此测量值记录为自动反转长度(ARL)。所有测量均由两名不同的研究人员进行两次。然后拔除牙齿并浸泡在20%的福尔马林溶液中48小时。固定后,将一根15号锉插入根管,从与Tri Auto ZX获得的相同参考点测量到根尖孔的实际根管长度,如在立体显微镜下所见。当锉尖在根管的解剖末端可见时,将其拔出0.5毫米并将此测量值记录为实际长度(AL)。所有测量值均以毫米表示,测量精度设定为0.5毫米。在5%置信水平下评估EL与ARL之间以及EL与AL测量值之间平均差异的显著性。

结果

在所有情况下,EL测量值与ARL一致。EL和ARL在10个(40%)根管中与AL一致,在其余15个根管(60%)中,AL测量值在14例中比EL和ARL长(+0.5毫米),在1例中短(-0.5毫米)。总体而言,AL比EL或ARL长,平均差异为-0.23毫米±0.32(P<0.05)。

结论

得出结论,Tri Auto ZX是有用且可靠的。Tri Auto ZX测量可防止过度预备。