A newly designed mass-detecting capillary viscometer is extended to measure the viscosity of whole blood over a range of shear rates without the use of anticoagulants in a clinical setting. In the present study as proof of principle, a single measurement of liquid-mass variation with time replaces the flow rate and pressure drop measurements that are usually required for the operation of a capillary tube viscometer. Using a load cell and capillary, we measured the change of mass flowing through capillary tube with respect to the time, m(t), from which viscosity and shear rate were mathematically calculated. For water and adulterated bloods, excellent agreement was found between the results from the mass-detecting capillary viscometer and those from a commercially available rotating viscometer. Also, the mass-detecting capillary viscometer measured the viscosity of unadulterated whole blood without heparin or EDTA. This new method overcomes the drawbacks of conventional viscometers in the measurement of the whole blood viscosity. First, the mass-detecting capillary viscometer can accurately and consistently measure the unadulterated blood viscosity over a range of shear rates in less than 2 min without any anticoagulants. Second, this design provides simplicity (i.e. ease of operation, no moving parts, and disposable) and low cost.
一种新设计的质量检测毛细管粘度计得到了扩展,可在临床环境中不使用抗凝剂的情况下,在一系列剪切速率范围内测量全血粘度。在本研究作为原理验证中,单次测量液体质量随时间的变化取代了毛细管粘度计操作通常所需的流速和压降测量。使用称重传感器和毛细管,我们测量了流经毛细管的质量随时间(m(t))的变化,并据此通过数学计算得出粘度和剪切速率。对于水和掺假血液,质量检测毛细管粘度计的结果与市售旋转粘度计的结果之间具有高度一致性。此外,质量检测毛细管粘度计可测量未掺假的全血粘度,无需使用肝素或乙二胺四乙酸(EDTA)。这种新方法克服了传统粘度计在测量全血粘度方面的缺点。首先,质量检测毛细管粘度计能够在不使用任何抗凝剂的情况下,在不到2分钟的时间内,在一系列剪切速率范围内准确且一致地测量未掺假血液的粘度。其次,这种设计具有操作简便(即易于操作、无活动部件且一次性使用)和成本低的特点。