National Blood Service, Brentwood, UK.
Vox Sang. 2002 Feb;82(2):76-9. doi: 10.1046/j.0042-9007.2001.00149.x.
To evaluate the HemoCue plasma haemoglobin (Hb) analyser for determining supernatant Hb in red cell concentrates (RCC) during storage and compare this to a spectrophotometric method.
Samples from RCC (n = 20) at days 0, 7, 21, 35 and 42 of storage were tested by both methods for supernatant Hb.
Results are given as median with range. There was a significant correlation (r = 0.97, P < 0.0001), but statistically significant difference between the two methods (0.72 (0.36-2.77) spectrophotometric vs. 0.70 (0.3-2.6) g/l HemoCue). However, the mean bias was negligible at 0.06 g/l. The HemoCue method gave a interassay coefficient of variation of 5.6% at 1.2 g/l and was linear to at least 20 g/l.
The HemoCue plasma haemoglobin analyser is a simple, rapid, reliable and accurate method for the determination of supernatant Hb in RCC at the end of storage and compares well to established methods for supernatant Hb measurement.
评估HemoCue血浆血红蛋白(Hb)分析仪在储存期间测定红细胞浓缩液(RCC)中上清液Hb的性能,并与分光光度法进行比较。
采用两种方法对储存第0、7、21、35和42天的RCC样本(n = 20)进行上清液Hb检测。
结果以中位数及范围表示。两种方法之间存在显著相关性(r = 0.97,P < 0.0001),但存在统计学显著差异(分光光度法为0.72(0.36 - 2.77)g/l,HemoCue法为0.70(0.3 - 2.6)g/l)。然而,平均偏差可忽略不计,为0.06 g/l。HemoCue法在1.2 g/l时批间变异系数为5.6%,线性范围至少达20 g/l。
HemoCue血浆血红蛋白分析仪是一种简单、快速、可靠且准确的方法,可用于测定储存末期RCC中的上清液Hb,与既定的上清液Hb测量方法相比效果良好。