Oh Michael Y, Abosch Aviva, Kim Seong H, Lang Anthony E, Lozano Andres M
Division of Neurosurgery, The Toronto Hospital, Western Division, University of Toronto, Toronto, Ontario, Canada.
Neurosurgery. 2002 Jun;50(6):1268-74; discussion 1274-6. doi: 10.1097/00006123-200206000-00017.
To determine the incidence of long-term hardware-related complications of deep brain stimulation (DBS).
The study design is a retrospective chart review of a single-surgeon, single-institution experience with DBS in 84 consecutive cases from 1993 to 1999. Only patients with a minimum follow-up of 1 year were considered. Five patients were excluded because trial stimulation failed to achieve pain relief (n = 4) or because the procedure was aborted owing to hemorrhage (n = 1). Seventy-nine patients received 124 permanent DBS electrode implants.
The mean follow-up period was 33 months, and the cumulative follow-up time was 217 patient-years or 310 electrode-years. Overall, 20 patients (25.3%) had 26 hardware-related complications involving 23 (18.5%) of the electrodes. There were 4 lead fractures, 4 lead migrations, 3 short or open circuits, 12 erosions and/or infections, 2 foreign body reactions, and one cerebrospinal fluid leak. The hardware-related complication rate per electrode-year was 8.4%. The most common complications were related to the electrode connectors. A significant finding was a high number of complications involving erosions or infections, which occurred in 7 of 12 instances as a late complication (beyond 12 mo).
Long-term follow-up reveals that hardware-related complications occur in a significant number of patients. Factors that lead to such complications must be identified and addressed to maximize the important benefits of DBS therapy.
确定深部脑刺激(DBS)与硬件相关的长期并发症的发生率。
本研究设计为对1993年至1999年期间一位外科医生在单一机构进行的84例连续DBS病例进行回顾性图表审查。仅考虑随访至少1年的患者。5例患者被排除,原因是试验性刺激未能缓解疼痛(n = 4)或因出血而中止手术(n = 1)。79例患者接受了124次永久性DBS电极植入。
平均随访期为33个月,累积随访时间为217患者年或310电极年。总体而言,20例患者(25.3%)发生了26例与硬件相关的并发症,涉及23根(18.5%)电极。有4例导线断裂、4例导线移位、3例短路或开路、12例侵蚀和/或感染、2例异物反应以及1例脑脊液漏。每电极年的硬件相关并发症发生率为8.4%。最常见的并发症与电极连接器有关。一个重要发现是大量并发症涉及侵蚀或感染,其中12例中有7例是晚期并发症(超过12个月)。
长期随访显示,相当数量的患者会发生与硬件相关的并发症。必须识别并解决导致此类并发症的因素,以最大限度地发挥DBS治疗的重要益处。