Bohnhorst B, Peter C S, Poets C F
Department of Neonatology and Pediatric Pulmonology, Hannover Medical School, Germany.
Arch Dis Child Fetal Neonatal Ed. 2002 Nov;87(3):F217-9. doi: 10.1136/fn.87.3.f217.
To determine the sensitivity and specificity of three newly developed pulse oximeters in the detection of hyperoxaemia, defined as an arterial partial pressure of oxygen (PaO(2)) of > 80 mm Hg.
SpO(2) readings from three oximeters (Agilent Viridia (AgV), Masimo SET (MaS), Nellcor Oxismart (NeO)) were documented in 56 infants (median gestational age at birth 35.5 weeks, range 24-41) whenever an arterial blood gas was taken for clinical purposes. Blood samples were analysed within one minute in a Radiometer ABL 505 blood gas analyser and OSM3 co-oximeter.
Between 280 and 291 blood gases were analysed for each instrument; 105-112 showed a PaO(2) > 80 mm Hg. At an upper alarm limit of 95%, the three instruments detected hyperoxaemia with 93-95% sensitivity. Specificity at this alarm level ranged from 26 to 45%. The mean (SD) difference between arterial oxygen saturation and SpO(2) (bias) was -0.25 (2.5)% for AgV, -0.06 (2.5)% for MaS, and -0.91 (2.6)% for NeO (p < 0.01, NeO v AgV and MaS).
These instruments detected hyperoxaemia with sufficient sensitivity at an upper alarm limit of 95%, but showed differences in their specificity, which was probably related to differences in measurement bias.
确定三种新开发的脉搏血氧仪检测高氧血症的敏感性和特异性,高氧血症定义为动脉血氧分压(PaO₂)>80 mmHg。
每当因临床目的采集动脉血气时,记录56例婴儿(出生时胎龄中位数为35.5周,范围24 - 41周)使用三种血氧仪(安捷伦Viridia(AgV)、Masimo SET(MaS)、Nellcor Oxismart(NeO))测得的SpO₂读数。血液样本在Radiometer ABL 505血气分析仪和OSM3血液一氧化碳测定仪中于1分钟内进行分析。
每种仪器分析了280至291份血气样本;105 - 112份样本显示PaO₂>80 mmHg。在上限报警值为95%时,三种仪器检测高氧血症的敏感性为93% - 95%。此报警水平下的特异性范围为26%至45%。AgV的动脉血氧饱和度与SpO₂之间的平均(标准差)差异(偏差)为 -0.25(2.5)%,MaS为 -0.06(2.5)%,NeO为 -0.91(2.6)%(p < 0.01,NeO与AgV和MaS相比)。
这些仪器在上限报警值为95%时检测高氧血症具有足够的敏感性,但特异性存在差异,这可能与测量偏差的差异有关。