Suppr超能文献

Temperature measurement on neurological pulse generators during MR scans.

作者信息

Kainz Wolfgang, Neubauer Georg, Uberbacher Richard, Alesch François, Chan Dulciana Dias

机构信息

Department of Mobile Communications Safety, ARC Seibersdorf Research, Austria.

出版信息

Biomed Eng Online. 2002 Sep 12;1:2. doi: 10.1186/1475-925x-1-2.

Abstract

According to manufacturers of both magnetic resonance imaging (MRI) machines, and implantable neurological pulse generators (IPGs), MRI is contraindicated for patients with IPGs. A major argument for this restriction is the risk to induce heat in the leads due to the electromagnetic field, which could be dangerous for the surrounding brain parenchyma. The temperature change on the surface of the case of an ITREL-III (Medtronic Inc., Minneapolis, MN) and the lead tip during MRI was determined. An anatomical realistic and a cubic phantom, filled with phantom material mimicking human tissue, and a typical lead configuration were used to imitate a patient who carries an IPG for deep brain stimulation. The measurements were performed in a 1.5 T and a 3.0 T MRI. 2.1 degrees C temperature increases at the lead tip uncovered the lead tip as the most critical part concerning heating problems in IPGs. Temperature increases in other locations were low compared to the one at the lead tip. The measured temperature increase of 2.1 degrees C can not be considered as harmful to the patient. Comparison with the results of other studies revealed the avoidance of loops as a practical method to reduce heating during MRI procedures.

摘要
https://cdn.ncbi.nlm.nih.gov/pmc/blobs/0adf/130994/605ce99f7e8c/1475-925X-1-2-1.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验