Biggs J T, Holland W H, Chang S, Hipps P P, Sherman W R
J Pharm Sci. 1976 Feb;65(2):261-8. doi: 10.1002/jps.2600650221.
A method for the measurement, in human plasma, of all tertiary and secondary tricyclic antidepressants prescribed in the United States is described. The method uses electron beam ionization GLC-mass spectrometry, employing a computer-controlled multiple-ion detector. This method, mass fragmentography, is used with internal standards for each drug. Plasma levels to as low as 10 ng/ml of the following drugs can be measured: amitriptyline, nortriptyline, doxepin, desmethyldoxepin, imipramine, desipramine, and protriptyline. Deuterium-labeled amitriptyline and imipramine are used as internal standards for those two drugs; for the other drugs, deuterated amitriptyline, nortriptyline, desmethyldoxepin, or desipramine is used. The method can measure up to 15 samples/hr, making it practical for large-scale studies of these drugs in patients. Spectra of each drug and examples of their analysis are given.
本文描述了一种用于测量美国所开处方中所有叔胺和仲胺三环类抗抑郁药在人体血浆中含量的方法。该方法采用电子束电离气相色谱 - 质谱联用技术,并使用计算机控制的多离子检测器。这种称为质量碎片分析法的方法与每种药物的内标物一起使用。可以测量以下药物低至10纳克/毫升的血浆水平:阿米替林、去甲替林、多塞平、去甲多塞平、丙咪嗪、地昔帕明和普罗替林。氘代阿米替林和丙咪嗪用作这两种药物的内标物;对于其他药物,则使用氘代阿米替林、去甲替林、去甲多塞平或地昔帕明。该方法每小时可测量多达15个样本,使其适用于对患者进行这些药物的大规模研究。给出了每种药物的光谱及其分析示例。