Design of an artificial lung compliance chamber for pulmonary replacement.

作者信息

Haft Jonathan W, Bull Joseph L, Rose Rebecca, Katsra Jeffrey, Grotberg James B, Bartlett Robert H, Hirschl Ronald B

机构信息

Department of Surgery, University of Michigan Health System, 1500 E. Medical Center Dr., Ann Arbor, Ml 48109, USA.

出版信息

ASAIO J. 2003 Jan-Feb;49(1):35-40. doi: 10.1097/00002480-200301000-00006.

Abstract

Matching the impedance of an artificial lung for pulmonary replacement to native pulmonary impedance is important in preventing right ventricular dysfunction. A lumped-parameter theoretical model and bench-top experiments were used to investigate the effect of a prototype compliance chamber on input impedance. The bench-top simulation consisted of a pulsatile flow generator, a prototype compliance chamber, and a low resistance artificial lung connected in series. Effective compliance was varied using pneumatic compression. The theoretical model considered a similar circuit with resistors before and after a compliance element. The bundle flow pulsatility (flow amplitude divided by average flow) and input impedance were calculated in the theoretical and experimental models. More compliance and lower upstream resistance result in lower bundle flow pulsatility and reduced first harmonic impedance. Matching the time scale of the circuit to the period of pulsatile flow also reduces impedance. The bench-top circuit demonstrated an optimal chamber pressure at which first harmonic impedance is reduced by 80%. The prototype compliance chamber in series with the artificial lung more closely matches native pulmonary impedance. The lumped-parameter model and the bench-top simulation will aid in the design and testing of compliance chamber modifications to improve its efficiency.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验