Suppr超能文献

食管假体置换

Prosthetic replacement of the esophagus.

作者信息

Salama F D

出版信息

J Thorac Cardiovasc Surg. 1975 Oct;70(4):739-46.

PMID:126337
Abstract

A tubular prosthesis was initially made of silicone rubber, spot-bonded on the inside and outside to a layer of knitted Dacron. This was covered on the outside by a loose layer of knitted Dacron, attached to the rubber only at the two ends, to allow incorporation by fibrous tissue and fixation of the prosthesis. Anastomoses were made with one layer of continuous sutures. There was initial incorporation of the graft. After 29 to 44 days the tube became loose and migrated, leaving a stricture. The prosthesis was redesigned with two loose layers of wider mesh. Incorporation occurred early with abundant fibrous reaction. Separation of the tube started at 40 days and was complete within 72 to 152 days. When the sutures became disengaged, bacteria entered the space between rubber and incorporated fabric, resulting in collagen lysis and separation of the tube. There were no anastomotic leaks in the entire series.

摘要

一种管状假体最初由硅橡胶制成,其内外通过点焊连接到一层针织涤纶上。外面覆盖有一层松散的针织涤纶,仅在两端与橡胶相连,以便纤维组织长入并固定假体。吻合采用一层连续缝合。移植物最初实现了长入。29至44天后,管子变得松动并移位,留下狭窄。假体重新设计为两层更宽网眼的松散结构。早期出现了长入,并伴有大量纤维反应。管子在40天开始分离,并在72至152天内完全分离。当缝线松开时,细菌进入橡胶与长入织物之间的间隙,导致胶原溶解和管子分离。整个系列中没有吻合口漏。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验