Thacker J G, Rodeheaver G, Moore J W, Kauzlarich J J, Kurtz L, Edgerton M T, Edlich R F
Am J Surg. 1975 Sep;130(3):374-80. doi: 10.1016/0002-9610(75)90408-0.
A comprehensive analysis of the mechanical performance of sutures has been made to provide information concerning the reliability and security of knotted sutures. The tests utilized in this analysis were designed to be easily reproduced by other investigators. The construction of the knot and the knot performance analysis were undertaken utilizing an Instron Tensile Tester. The mechanical reliability of each knotted suture was determined by measuring the number of throws to reach knot break, the expected slippage of the knot when it reaches knot break, and the maximal holding force at knot break. On the basis of these measurements, recommendations are made for the use of a suture at operation.