Tehrani Mana, Krummenauer Frank, Kumar Rajiv, Dick H Burkhard
Department of Ophthalmology, Johannes-Gutenberg University, Mainz, Germany.
J Cataract Refract Surg. 2003 Apr;29(4):747-52. doi: 10.1016/s0886-3350(02)01739-x.
To compare and contrast axial length (AL) measurements assessed by ultrasound biometry and optical biometry.
Department of Ophthalmology, Johannes Gutenberg-University, Mainz, Germany.
Optical biometry and ultrasound biometry were performed in 360 eyes to measure AL.
In eyes with a clear-appearing crystalline lens, there was a statistically significant median difference in AL measurements between the 2 methods in eyes with a normal or long length. In eyes with cataract, there was a statistically significant median difference in AL measurements in eyes with a normal or short length. In these cases, optical biometry produced larger readings than ultrasound biometry. In eyes with a posterior chamber intraocular lens, there was no statistically significant difference between the 2 methods in eyes of all lengths. In all eyes, the 2 methods produced reliable measurements.
These results show that optical biometry and ultrasound applanation biometry give statistically significant differences in AL measurement in patients with cataract and normal lenses. In these cases, optical biometry provided clinically relevant larger values than ultrasound applanation. Optical biometry provides an alternative to ultrasound biometry. However, ultrasound biometry may be indicated in certain complex cases.
比较和对比超声生物测量法与光学生物测量法所测得的眼轴长度(AL)。
德国美因茨约翰内斯·古腾堡大学眼科。
对360只眼睛进行光学生物测量法和超声生物测量法以测量眼轴长度。
在晶状体外观清晰的眼睛中,两种测量方法在正常或较长眼轴长度的眼睛中测得的眼轴长度中位数存在统计学显著差异。在患有白内障的眼睛中,两种测量方法在正常或较短眼轴长度的眼睛中测得的眼轴长度中位数存在统计学显著差异。在这些情况下,光学生物测量法测得的数值大于超声生物测量法。在植入后房型人工晶状体的眼睛中,两种测量方法在各种眼轴长度的眼睛中均无统计学显著差异。在所有眼睛中,两种测量方法均能提供可靠的测量结果。
这些结果表明,光学生物测量法与超声压平生物测量法在白内障患者和晶状体正常的患者的眼轴长度测量中存在统计学显著差异。在这些情况下,光学生物测量法提供的临床相关数值大于超声压平测量法。光学生物测量法为超声生物测量法提供了一种替代方法。然而,在某些复杂情况下可能仍需使用超声生物测量法。