Prevention of tracheal aspiration using the pressure-limited tracheal tube cuff.

作者信息

Young P J, Basson C, Hamilton D, Ridley S A

机构信息

Intensive Care Unit, Norfolk and Norwich Hospital, Norwich, UK.

出版信息

Anaesthesia. 1999 Jun;54(6):559-63. doi: 10.1046/j.1365-2044.1999.00850.x.

Abstract

A new design of tracheal tube cuff, the pressure-limited cuff, used with a constant-pressure inflation system, was compared with a high-volume low-pressure cuffed tracheal tube for leakage of dye placed in the subglottic space into the trachea. Patients requiring ventilation on the intensive care unit were randomly allocated into two groups, one for each type of cuff, and blue food dye was instilled daily via a fine catheter above the cuff into the subglottic space. There were eight patients in the high-volume low-pressure group and seven in the pressure-limited cuff group. Dye leaked into the trachea in seven (87%) of the high-volume low-pressure group compared with none (0%) of the pressure-limited cuff group (p < 0.01). This study demonstrates that the pressure-limited cuffed tracheal tube, in combination with a constant-pressure inflation device, prevents leakage of fluid into the lungs that occurs with high-volume low-pressure cuffs in the critically ill, intubated patient.

摘要

一种新型气管导管套囊——压力限制套囊,与恒压充气系统配合使用,与大容量低压套囊气管导管相比,用于检测声门下间隙的染料漏入气管的情况。在重症监护病房需要通气的患者被随机分为两组,每组使用一种类型的套囊,每天通过细导管将蓝色食用染料注入套囊上方的声门下间隙。大容量低压组有8名患者,压力限制套囊组有7名患者。大容量低压组有7名(87%)患者出现染料漏入气管,而压力限制套囊组无一例(0%)出现这种情况(p<0.01)。本研究表明,压力限制套囊气管导管与恒压充气装置联合使用,可防止重症插管患者使用大容量低压套囊时出现的液体漏入肺部的情况。