Suppr超能文献

[The indirect measurement of arterial pressure as a function of cuff width in patients in the immediate preoperative phase, the phase of entry into the operating room and in the postanesthesia phase].

作者信息

Peniche A de C, Arcuri E A

出版信息

Rev Esc Enferm USP. 1992 Aug;26(2):243-56. doi: 10.1590/0080-6234199202600200243.

Abstract

The goal of this study was to compare blood pressure measures using two sizes of cuffs: one standard width (12 cm) and other with correct width that is 0.38 of arm circumference as recommended by American Heart Association. The comparisons were done among surgical patients in mediate perioperative phase, reception area of operation room and postoperative phase during the staying of the patient in the recovery room. The results demonstrated that the arterial blood pressure was hypoestimated by the use of the standard width cuff, reaching values up 30 mmHg in the systolic arterial pressure and 30 mmHg in the diastolic ones.