Cook S D, Baffes G C, Palafox A J, Wolfe M W, Burgess A
Department of Orthopaedic Surgery, Tulane University School of Medicine, New Orleans, LA 70112.
J Oral Implantol. 1992;18(4):354-65.
Hydroxylapatite (HA)-coated and grit-blasted (non-HA-coated) titanium dental implants were inserted into healed extraction sites of canine mandibles. After six weeks, the animals were killed and the implants mechanically tested in torsion to failure. Interface attachment strength, implant/tissue compatibility, integrity of the HA coating, and the location of interface failure were evaluated. Mechanical testing demonstrated an interface torsional strength of 3.98 +/- 0.93 MPa for the HA-coated implants and 2.25 +/- 0.65 MPa for the grit-blasted implants. This represents a 76.9% improvement in the maximum torsional interface strength, and is statistically-significant (p = 0.0004). On qualitative histologic analysis, interface failure was seen to occur primarily at the HA/implant interface, although failure through the HA coating and regions of bone/HA interface failure were observed. The HA-coated implants had bone in direct apposition to their surface with no fibrous tissue interposition. The grit-blasted implants also had regions of direct bone-implant apposition, but these areas were limited to a smaller proportion of the total interface area. There was no evidence of breakdown or change in thickness of the HA coating.
将羟基磷灰石(HA)涂层和喷砂处理(非HA涂层)的钛制牙科植入物植入犬下颌骨愈合的拔牙部位。六周后,处死动物,并对植入物进行扭转至失效的力学测试。评估界面附着强度、植入物/组织相容性、HA涂层的完整性以及界面失效的位置。力学测试表明,HA涂层植入物的界面抗扭强度为3.98±0.93MPa,喷砂处理植入物的界面抗扭强度为2.25±0.65MPa。这表明最大扭转界面强度提高了76.9%,具有统计学意义(p = 0.0004)。在定性组织学分析中,尽管观察到HA涂层贯穿失效以及骨/HA界面区域失效,但界面失效主要发生在HA/植入物界面。HA涂层植入物表面有骨直接附着,无纤维组织插入。喷砂处理的植入物也有骨与植入物直接附着的区域,但这些区域在总界面面积中所占比例较小。没有证据表明HA涂层有破裂或厚度变化。