Reljic R, Ries M, Anić N, Ries B
Faculty of Food Technology and Biotechnology, Laboratory of Biochemistry, Zagreb, Croatia.
Clin Chem. 1992 Apr;38(4):522-5.
We describe a colorimetric assay for glucose determination in human serum, with use of the chromogen 2-amino-4-hydroxybenzenesulfonic acid (AHBS), glucose oxidase, and peroxidase. With this assay, glucose concentrations less than or equal to 27.8 mmol/L can be measured in serum, with a sample/reagent volume ratio as low as 0.0025. The chromogen itself is easily soluble in water and does not require other components for the color change, making the reagent composition less complex. A single working reagent is used, and the reaction is completed within 10 min at 37 degrees C. The absorbance of the yellow reaction product is measured at 415 nm, and a blank sample measurement is not needed. The average analytical recovery of glucose in different human sera was 97.6%, with no significant interference of reducing compounds in serum. The results of the recommended procedure correlated well with those of the phenol/4-aminoantipyrine method of Trinder.
我们描述了一种用于测定人血清中葡萄糖的比色测定法,该方法使用显色剂2-氨基-4-羟基苯磺酸(AHBS)、葡萄糖氧化酶和过氧化物酶。采用这种测定法,血清中葡萄糖浓度小于或等于27.8 mmol/L时即可测定,样品/试剂体积比低至0.0025。该显色剂本身易溶于水,且颜色变化不需要其他成分,使得试剂组成不那么复杂。使用单一工作试剂,在37℃下10分钟内即可完成反应。在415nm处测量黄色反应产物的吸光度,无需测量空白样品。不同人血清中葡萄糖的平均分析回收率为97.6%,血清中的还原性化合物无明显干扰。推荐方法的结果与Trinder的酚/4-氨基安替比林法的结果相关性良好。