Suppr超能文献

热膜风速仪——一种测定血流速度的方法。I. 与电磁血流计的体外比较。

The hot-film anemometer--a method for blood velocity determination. I. In vitro comparison with the electromagnetic blood flowmeter.

作者信息

Paulsen P K

出版信息

Eur Surg Res. 1980;12(2):140-8. doi: 10.1159/000128119.

Abstract

In an invitro flowmodel a constrant temperature hot-film anemometer was tested, using conical needle and catheter-mounted probes. Calibration, linearization and zero-point determination as well as sensitivity change with hematocrit, angulation, flow direction and dirt deposits on the film are described. Curves were compared with those obtained simultaneously from an electromagnetic flowmeter. The hog-film anemometer was direction-insensitive and signals were seen 0.01--0.03 sec before the flowmeter signals. The frequency response of the anemometer was sufficiently higher to register turbulent flow. In 27 simultaneous measurements the mean anemometer results were 6 +/- 8% (+/- SD) higher than the flowmeter results and the peak results correspondingly 16 +/- 6%. Both differences was significant (p < 0.001). Qualitatively hog-film anemometer curves were comparable to electromagnetic flowmeter curves. However, certain differences were demonstrated.

摘要

在一个体外血流模型中,使用锥形针和导管安装探头对恒温热膜风速仪进行了测试。描述了校准、线性化、零点测定以及灵敏度随血细胞比容、角度、血流方向和膜上污垢沉积的变化。将曲线与同时从电磁流量计获得的曲线进行了比较。热膜风速仪对方向不敏感,信号比流量计信号提前0.01 - 0.03秒出现。风速仪的频率响应足够高,能够记录湍流。在27次同步测量中,风速仪的平均结果比流量计结果高6±8%(±标准差),峰值结果相应地高16±6%。这两个差异均具有显著性(p < 0.001)。定性地说,热膜风速仪曲线与电磁流量计曲线具有可比性。然而,也显示出了某些差异。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验