Paulsen P K
Eur Surg Res. 1980;12(2):140-8. doi: 10.1159/000128119.
In an invitro flowmodel a constrant temperature hot-film anemometer was tested, using conical needle and catheter-mounted probes. Calibration, linearization and zero-point determination as well as sensitivity change with hematocrit, angulation, flow direction and dirt deposits on the film are described. Curves were compared with those obtained simultaneously from an electromagnetic flowmeter. The hog-film anemometer was direction-insensitive and signals were seen 0.01--0.03 sec before the flowmeter signals. The frequency response of the anemometer was sufficiently higher to register turbulent flow. In 27 simultaneous measurements the mean anemometer results were 6 +/- 8% (+/- SD) higher than the flowmeter results and the peak results correspondingly 16 +/- 6%. Both differences was significant (p < 0.001). Qualitatively hog-film anemometer curves were comparable to electromagnetic flowmeter curves. However, certain differences were demonstrated.
在一个体外血流模型中,使用锥形针和导管安装探头对恒温热膜风速仪进行了测试。描述了校准、线性化、零点测定以及灵敏度随血细胞比容、角度、血流方向和膜上污垢沉积的变化。将曲线与同时从电磁流量计获得的曲线进行了比较。热膜风速仪对方向不敏感,信号比流量计信号提前0.01 - 0.03秒出现。风速仪的频率响应足够高,能够记录湍流。在27次同步测量中,风速仪的平均结果比流量计结果高6±8%(±标准差),峰值结果相应地高16±6%。这两个差异均具有显著性(p < 0.001)。定性地说,热膜风速仪曲线与电磁流量计曲线具有可比性。然而,也显示出了某些差异。