Frayne R, Holdsworth D W, Gowman L M, Rickey D W, Drangova M, Fenster A, Rutt B K
Department of Medical Biophysics, University Hospital, University of Western Ontario, London, Canada.
J Magn Reson Imaging. 1992 Sep-Oct;2(5):605-12. doi: 10.1002/jmri.1880020522.
A novel computer-controlled flow simulator for use in magnetic resonance (MR) flow experiments was evaluated. The accuracy in constant-flow mode was better than 1%. The accuracy in pulsatile-flow mode was found to be dependent on the interconnecting tubing. The short-term and long-term reproducibilities of pulsatile waveforms were less than or equal to 0.4 mL/sec (1 standard deviation). Increased response times due to the lengths of tubing required in MR flow experiments were surmounted by using a modified tubing configuration and precompensated waveforms. Piston reversal was found not to cause major difficulties in MR flow experiments.