In vitro effects of light on serum bilirubin subfractions measured by high-performance liquid chromatography: comparison with four routine methods.

作者信息

Ihara H, Nakamura H, Aoki Y, Aoki T, Yoshida M

机构信息

Department of Laboratory Medicine, Toho University, Tokyo, Japan.

出版信息

Clin Chem. 1992 Oct;38(10):2124-9.

PMID:1395005
Abstract

The effects of light on serum bilirubin subfractions in vitro were investigated by HPLC and four routine methods for bilirubin analysis. By HPLC, the rate of photodegradation of unconjugated bilirubin (Bu) was nearly twice that of monoconjugated bilirubin (mBc) and threefold that of diconjugated bilirubin (dBc); delta bilirubin (Bd) was most stable against photoirradiation. In the diazo method, the rate of photodegradation of direct bilirubin was almost the same as that of the sum of mBc, dBc, and Bd determined by the HPLC method. However, the rate of photodegradation of indirect bilirubin was significantly lower (P < 0.001) than that obtained by HPLC, because approximately 30% of the bilirubin photoproducts reacted with the diazo reagent as indirect bilirubin. The rate of photodegradation of total bilirubin determined by the direct spectrometric method was lower than that determined by the diazo method, but equal to that of the total peak areas of HPLC. In the Ektachem method, bilirubin photoproducts affected total bilirubin negligibly, and Bc and Bu positively, so that the value of Bd decreased. In the bilirubin oxidase method, bilirubin photoproducts were oxidized enzymatically by both the total and direct bilirubin reagents. We re-emphasize the importance of shielding serum from light to avoid generating bilirubin photoproducts that interfere with the accurate determination of serum bilirubin subfractions. We also recommend HPLC analysis as a standard method for bilirubin measurement.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验