Chester A E, Dorr A E, Lund K R, Wood L D
Institute of Toxicologic Sciences, Syntex Research, Palo Alto, California 94303.
Fundam Appl Toxicol. 1992 Jul;19(1):64-8. doi: 10.1016/0272-0590(92)90029-h.
Systolic (SP), diastolic (DP), and mean arterial blood pressures (MAP) and pulse rate (PR) were recorded on treated and untreated conscious cynomolgus monkeys by the oscillometric method (Dinamap 1846SX/P). Each monkey was placed in a restraining tube with the cuff placed on the base on the shaved tail. Measurements were taken on untreated animals once or twice a day for 4 weeks. The mean and standard deviations for SP, DP, MAP, and PR were approximately 121 +/- 17, 60 +/- 14, and 84 +/- 17 mm Hg and 193 +/- 18 pulses/min, respectively. One male and one female cynomolgus monkey were treated with isoproterenol, norepinephrine, and nitroprusside. Blood pressure was measured indirectly with the cuff on the tail and directly with an indwelling catheter in the descending thoracic aorta. Although the oscillometric method was not as sensitive as the catheter, the oscillometric method detected a change in the same direction of SP, DP, MAP, and PR compared with the direct method for all drugs administered. The monitor was reliable and sufficiently accurate to conclude that it may be useful in toxicology studies for evaluation of blood pressure parameters in conscious cynomolgus monkeys.
采用示波法(Dinamap 1846SX/P)记录经治疗和未经治疗的清醒食蟹猴的收缩压(SP)、舒张压(DP)、平均动脉压(MAP)和脉搏率(PR)。将每只猴子置于约束管中,袖带置于剃毛后的尾巴根部。对未经治疗的动物每天测量1次或2次,共测量4周。SP、DP、MAP和PR的平均值及标准差分别约为121±17、60±14、84±17 mmHg和193±18次/分钟。对1只雄性和1只雌性食蟹猴给予异丙肾上腺素、去甲肾上腺素和硝普钠治疗。通过尾巴上的袖带间接测量血压,并通过降主动脉内的留置导管直接测量血压。虽然示波法不如导管法敏感,但与直接测量法相比,示波法检测到SP、DP、MAP和PR在所有给药药物作用下的变化方向相同。该监测仪可靠且足够准确,可得出结论,其在毒理学研究中评估清醒食蟹猴的血压参数可能有用。